A NEW METHOD OF MEASURING CARDIAC-OUTPUT IN MAN USING LITHIUM DILUTION

被引:98
作者
LINTON, RAF [1 ]
BAND, DM [1 ]
HAIRE, KM [1 ]
机构
[1] UNITED MED & DENT SCH GUYS & ST THOMAS HOSP,ST THOMAS HOSP,SHERRINGTON SCH PHYSIOL,LONDON SE1 7EH,ENGLAND
关键词
HEART; CARDIAC OUTPUT; MEASUREMENT TECHNIQUES; THERMODILUTION;
D O I
10.1093/bja/71.2.262
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We describe a new indicator dilution method of measuring cardiac output in man. A bolus injection of lithium chloride 0.6 mmol was given via a central venous catheter and arterial plasma [Li+] recorded using a specially developed sensor incorporating an Li+-selective electrode. Cardiac output was derived from the lithium dilution curve, with a correction for packed cell volume. Lithium dilution cardiac output (LiDCO) was compared with thermodilution cardiac output (TD) using 22 lithium sensors in nine patients. For each sensor, one LiDCO was measured immediately before and one immediately after three TD estimations and mean values of LiDCO and TD derived. The correlation coefficient, r, was 0.89; slope of the regression 0.84; y intercept 0.72; bias 0.3 (0.5) litre min-1 (mean (TD-LiDCO) (1 SD). LiDCO appeared to be a safe, simple and accurate technique which does not require insertion of a pulmonary artery catheter.
引用
收藏
页码:262 / 266
页数:5
相关论文
共 15 条
[1]  
BAND DM, 1986, J PHYSIOL-LONDON, V372, pP2
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   MEASUREMENT OF CARDIAC OUTPUT BY THERMAL DILUTION IN MAN [J].
BRANTHWAITE, MA ;
BRADLEY, RD .
JOURNAL OF APPLIED PHYSIOLOGY, 1968, 24 (03) :434-+
[4]   ANALYSIS OF ERRORS IN MEASUREMENT OF CARDIAC-OUTPUT BY SIMULTANEOUS DYE AND THERMAL DILUTION IN CARDIOTHORACIC SURGICAL PATIENTS [J].
FISCHER, AP ;
BENIS, AM ;
JURADO, RA ;
SEELY, E ;
TEIRSTEIN, P ;
LITWAK, RS .
CARDIOVASCULAR RESEARCH, 1978, 12 (03) :190-199
[5]  
HAUSEN B, 1992, ANESTH ANALG, V74, P800
[6]   INACCURACY OF CARDIAC-OUTPUT BY THERMODILUTION DURING ACUTE TRICUSPID REGURGITATION [J].
HEERDT, PM ;
POND, CG ;
BLESSIOS, GA ;
ROSENBLOOM, M .
ANNALS OF THORACIC SURGERY, 1992, 53 (04) :706-708
[7]   LITHIUM ION-SELECTIVE ELECTRODES BASED ON CROWN ETHERS FOR SERUM LITHIUM ASSAY [J].
KIMURA, K ;
OISHI, H ;
MIURA, T ;
SHONO, T .
ANALYTICAL CHEMISTRY, 1987, 59 (19) :2331-2334
[8]  
PLENGE P, 1988, LITHIUM INORGANIC PH, P135
[9]   TRANSESOPHAGEAL PULSED WAVE DOPPLER MEASUREMENT OF CARDIAC-OUTPUT DURING MAJOR VASCULAR-SURGERY - COMPARISON WITH THE THERMODILUTION TECHNIQUE [J].
RYAN, T ;
PAGE, R ;
BOUCHIERHAYES, D ;
CUNNINGHAM, AJ .
BRITISH JOURNAL OF ANAESTHESIA, 1992, 69 (01) :101-104
[10]  
SHELLEY RK, 1988, LITHIUM INORGANIC PH, P115