Protocol for Evaluation of Robotic Technology in Orthopedic Surgery

被引:2
作者
Masjedi, Milad [1 ]
Jaffry, Zahra [1 ]
Harris, Simon [1 ]
Cobb, Justin [1 ]
机构
[1] Imperial Coll London, Charing Cross Hosp, MSk Lab, London W6 8RF, England
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
D O I
10.1155/2013/194683
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In recent years, robots have become commonplace in surgical procedures due to their high accuracy and repeatability. The Acrobot Sculptor is an example of such a robot that can assist with unicompartmental knee replacement. In this study, we aim to evaluate the accuracy of the robot (software and hardware) in a clinical setting. We looked at (1) segmentation by comparing the segmented data from Sculptor software to other commercial software, (2) registration by checking the inter-and intraobserver repeatability of selecting set points, and finally (3) sculpting (n = 9 cases) by evaluating the achieved implant position and orientation relative to that planned. The results from segmentation and registration were found to be accurate. The highest error was observed in flexion extension orientation of femoral implant (0.4 +/- 3.7 degrees). Mean compound rotational and translational errors for both components were 2.1 +/- 0.6mmand 3 +/- 0.8 degrees for tibia and 2.4 +/- 1.2mmand 4.3 +/- 1.4 degrees for the femur. The results from all processes used in Acrobot were small. Validation of robot in clinical settings is highly vital to ensure a good outcome for patients. It is therefore recommended to follow the protocol used here on other available similar products.
引用
收藏
页数:5
相关论文
共 24 条
[1]  
Aggarwal R, 2004, Int J Surg, V2, P106, DOI 10.1016/S1743-9191(06)60055-1
[2]   Imageless versus image-based registration in navigated arthroscopy of the hip A CADAVER-BASED ASSESSMENT [J].
Audenaert, E. ;
Smet, B. ;
Pattyn, C. ;
Khanduja, V. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (05) :624-629
[3]   Active Shape Modeling of the Hip in the Prediction of Incident Hip Fracture [J].
Baker-LePain, Julie C. ;
Luker, Kali R. ;
Lynch, John A. ;
Parimi, Neeta ;
Nevitt, Michael C. ;
Lane, Nancy E. .
JOURNAL OF BONE AND MINERAL RESEARCH, 2011, 26 (03) :468-474
[4]   Robots in orthopaedic surgery - Past, present, and future [J].
Bargar, William L. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (463) :31-36
[5]  
Biomet, 2012, OXF PART KNEE MAN SU
[6]   Hands-on robotic unicompartmental knee replacement - A prospective, randomised controlled study of the Acrobot system [J].
Cobb, J ;
Henckel, J ;
Gomes, P ;
Harris, S ;
Jakopec, M ;
Rodriguez y Baena, F ;
Barrett, A ;
Davies, B .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (02) :188-197
[7]   Robotic control in knee joint replacement surgery [J].
Davies, B. L. ;
Rodriguez y Baena, F. M. ;
Barrett, A. R. W. ;
Gomes, M. P. S. F. ;
Harris, S. J. ;
Jakopec, M. ;
Cobb, J. P. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H1) :71-80
[8]   Accuracy of Dynamic Tactile-Guided Unicompartmental Knee Arthroplasty [J].
Dunbar, Nicholas J. ;
Roche, Martin W. ;
Park, Brian H. ;
Branch, Sharon H. ;
Conditt, Michael A. ;
Banks, Scott A. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (05) :803-808
[9]   The role of registration in accurate surgical guidance [J].
Fitzpatrick, J. M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H5) :607-622
[10]  
Hananouchi T, 2008, COMPUT AIDED SURG, V13, P23, DOI [10.1080/10929080701882580, 10.3109/10929080701882580]