CLINICAL-EVALUATION OF CAP SYSTEM AND RAST IN THE MEASUREMENT OF SPECIFIC IGE

被引:27
作者
PASTORELLO, EA
INCORVAIA, C
PRAVETTONI, V
MARELLI, A
FARIOLI, L
GHEZZI, M
机构
[1] PHARMACIA SPA,DIV DIAGNOST,MILAN,ITALY
[2] UNIV MILAN,INST INTERNAL MED INFECT DIS & IMMUNOPATHOL,DEPT INTERNAL MED 1,I-20122 MILAN,ITALY
关键词
D O I
10.1111/j.1398-9995.1992.tb00665.x
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
We investigated the diagnostic value of a new in vitro test, Pharmacia CAP System (Pharmacia Diagnostics AB, Uppsala, Sweden), for the quantitative measurement of allergen-specific IgE antibodies by comparison with RAST in 2 groups of patients, 71 atopic and 48 non-atopic. In the last 20 years RAST has supplied a good diagnostic tool, but this test presents some problems, the main one being sensitivity. The new test has a solid phase able to bind even very small amounts of specific IgE and an anti-IgE tracer with very low cross-reactivity with other immunoglobulins, thus presenting more favourable conditions. From the analysis of our results, Pharmacia CAP System gave higher sensitivity (94% compared to 88% of RAST) with no loss of specificity (96% for both tests). The reliability of these results is ensured by the proper selection of patients who were all suffering from pollinosis and were clinically diagnosed as certainly hypersensitive to a single pollen. A positive trend was found between severity of asthma and levels of specific IgE for timothy. Pharmacia CAP System appears to identify a larger number of atopic patients than RAST.
引用
收藏
页码:463 / 466
页数:4
相关论文
共 18 条