PLATELET AND COAGULATION-FACTOR VARIATIONS INDUCED IN-VITRO BY POLYETHYLENE TEREPHTHALATE (DACRON(R)) COATED WITH PYROLYTIC CARBON

被引:29
作者
CENNI, E
ARCIOLA, CR
CIAPETTI, G
GRANCHI, D
SAVARINO, L
STEA, S
CAVEDAGNA, D
CURTI, T
FALSONE, G
PIZZOFERRATO, A
机构
[1] UNIV BOLOGNA,CATTEDRA CHIRURG VASC,I-40126 BOLOGNA,ITALY
[2] USL 27,CTR REG SERV TRANFUS,BOLOGNA,ITALY
关键词
PYROLYTIC CARBON; PLATELETS; INTRINSIC COAGULATION PATHWAY;
D O I
10.1016/0142-9612(95)94903-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The haemocompatibility of polyethylene terephthalate (Dacron (R)) coated with pyrolytic carbon was examined in vitro, evaluating its capability of inducing adhesion and platelet activation, and of modifying the intrinsic coagulation pathway. Platelet adhesion was evaluated by counting platelets before and after in vitro contact of human plasma with the material under examination. Platelet activation was evaluated by determining platelet factor 4 (PF4) and thromboxane B-2. intrinsic coagulation pathway alterations were studied by determining activated partial thromboplastin time (APTT) and the activity of single factors. The results obtained show that pyrolytic carbon-coated Dacron induces platelet adhesion, reduction in platelet volume and lower increase in thromboxane production than that obtained after contact with uncoated Dacron. Pyrolytic carbon-coated Dacron does not induce PF4 release, contrary to uncoated Dacron induces a significant release. Moreover, pyrolytic carbon-coated Dacron, induces a significant extension of APTT by reducing the activity of intrinsic pathway factors, particularly factor XI.
引用
收藏
页码:973 / 976
页数:4
相关论文
共 17 条
  • [1] TISSUE REACTION TO FABRICS COATED WITH TURBOSTRATIC CARBON - SUBCUTANEOUS VERSUS VASCULAR IMPLANTS
    AEBISCHER, P
    GODDARD, MB
    SASKEN, HF
    HUNTER, TJ
    GALLETTI, PM
    [J]. BIOMATERIALS, 1988, 9 (01) : 80 - 85
  • [2] HEPARINIZABLE MATERIALS .4. SURFACE-GRAFTING ON POLY(ETHYLENE-TEREPHTHALATE) OF HEPARIN-COMPLEXING POLY(AMIDO-AMINE) CHAINS
    BARBUCCI, R
    BENVENUTI, M
    CASINI, G
    FERRUTI, P
    TEMPESTI, F
    [J]. BIOMATERIALS, 1985, 6 (02) : 102 - 104
  • [3] PYROLYTIC CARBONS DEPOSITED IN FLUIDIZED BEDS AT 1200 TO 1400DEGREESC FROM VARIOUS HYDROCARBONS
    BARD, RJ
    BAXMAN, HR
    BERTINO, JP
    OROURKE, JA
    [J]. CARBON, 1968, 6 (05) : 603 - &
  • [4] BOKROS JC, 1977, CARBON, V15, P355, DOI 10.1016/0008-6223(77)90324-4
  • [5] CENNI E, 1992, ADV BIOMAT, V10, P235
  • [6] CENNI E, IN PRESS J BIOMED MA
  • [7] SURFACE ATOMIC AND DOMAIN-STRUCTURES OF BIOMEDICAL CARBONS OBSERVED BY SCANNING TUNNELING MICROSCOPY (STM)
    FENG, L
    ANDRADE, JD
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (02): : 177 - 182
  • [8] FENG L, 1994, 20TH ANN M SOC BIOM, P249
  • [9] Granstrom E, 1978, Adv Prostaglandin Thromboxane Res, V5, P119
  • [10] Haubold A.D., 1983, ASAIO J, V6, P88