Robust control of depth of anesthesia

被引:114
作者
Dumont, Guy A. [1 ]
Martinez, Arturo [1 ]
Ansermino, J. Mark [2 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V6T 1Z4, Canada
[2] Univ British Columbia, Dept Anesthesia Pharmacol & Therapeut, Vancouver, BC V6T 1Z4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
automated drug delivery; anesthesia; robust control; CLOSED-LOOP CONTROL; BISPECTRAL INDEX; PERFORMANCE ASSESSMENT; PROPOFOL; DELIVERY; MULTICENTER; SIMULATION; SYSTEM;
D O I
10.1002/acs.1087
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents a systematic procedure to design both robust PID controllers and robust controllers based on fractional calculus (based on Commande Robuste d'Ordre Non Entier, or CRONE methodology) to regulate the hypnotic state of anesthesia with the intravenous administration of propofol. The objective of the controllers is to provide an adequate drug administration regimen for propofol to avoid under or over dosing of the patients. The controllers are designed to compensate for the patients inherent drug-response variability (uncertainty), to achieve good Output disturbance rejection, and to attain good set point response. The performance of the controllers is assessed by calculating typical time domain measures and using the median PE, median absolute PE, divergence, and wobble. Copyright (C) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:435 / 454
页数:20
相关论文
共 19 条
  • [1] Closed-loop control of anesthesia using bispectral index - Performance assessment in patients undergoing major orthopedic surgey under combined general and regional anesthesia
    Absalom, AR
    Sutcliffe, N
    Kenny, GN
    [J]. ANESTHESIOLOGY, 2002, 96 (01) : 67 - 73
  • [2] Introduction to automated drug delivery in clinical anesthesia
    Bibian, S
    Ries, CR
    Huzmezan, M
    Dumont, G
    [J]. EUROPEAN JOURNAL OF CONTROL, 2005, 11 (06) : 535 - 557
  • [3] Estimation of optimal modeling weights for a Bayesian-based closed-loop system for propofol administration using the bispectral index as a controlled variable: A simulation study
    De Smet, Tom
    Struys, Michel M. R. F.
    Greenwald, Scott
    Mortier, Eric P.
    Shafer, Steven L.
    [J]. ANESTHESIA AND ANALGESIA, 2007, 105 (06) : 1629 - 1638
  • [4] Modeling and closed-loop control of hypnosis by means of bispectral index (BIS) with isoflurane
    Gentilini, A
    Rossoni-Gerosa, M
    Frei, CW
    Wymann, R
    Morari, M
    Zbinden, AM
    Schnider, TW
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (08) : 874 - 889
  • [5] Automated anesthesia - Fact or fantasy?
    Glass, PSA
    Rampil, IJ
    [J]. ANESTHESIOLOGY, 2001, 95 (01) : 1 - 2
  • [6] Development and clinical application of electroencephalographic bispectrum monitoring
    Johansen, JW
    Sebel, PS
    [J]. ANESTHESIOLOGY, 2000, 93 (05) : 1336 - 1344
  • [7] Titration of propofol for anesthetic induction and maintenance guided by the bispectral index:: Closed-loop versus manual control -: A prospective, randomized, multicenter study
    Liu, N
    Chazot, T
    Genty, A
    Landais, A
    Restoux, A
    McGee, K
    Laloë, PA
    Trillat, B
    Barvais, L
    Fischler, M
    [J]. ANESTHESIOLOGY, 2006, 104 (04) : 686 - 695
  • [8] Fractional-Order Control: Theory and Applications in Motion Control
    Ma, Chengbin
    Hori, Yoichi
    [J]. IEEE INDUSTRIAL ELECTRONICS MAGAZINE, 2007, 1 (04) : 6 - 16
  • [9] ELECTROENCEPHALOGRAPHICALLY CONTROLLED ANESTHESIA IN ABDOMINAL SURGERY
    MAYO, CW
    BICKFORD, RG
    FAULCONER, A
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1950, 144 (13): : 1081 - 1083
  • [10] Hemodynamic control using direct model reference adaptive control - Experimental results
    Palerm, CC