REAL-TIME ABDOMINAL FETAL ECG RECORDING USING A HARDWARE CORRELATOR

被引:24
作者
ABBOUD, S
ALALUF, A
EINAV, S
SADEH, D
机构
[1] Biomedical Engineering Program, Faculty of Engineering, Tel Aviv University, Tel Aviv
关键词
FETAL MONITORING; ABDOMINAL ECG; SIGNAL AVERAGING;
D O I
10.1016/0010-4825(92)90021-E
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
A real-time fetal ECG monitoring system using abdominal recording is presented. The system is based on an IBM AT compatible personal computer. The computer lacks the performance required for real-time analysis. Therefore, a new design of a fast hardware correlator board was developed to enhance the computer throughput. The technique is based on a cross-correlation procedure. An averaged maternal ECG waveform is derived using the cross-correlation function for the waveform's alignment. With this procedure a template signal corresponding to one complete maternal ECG is obtained. The averaged maternal ECG is then subtracted from the abdominal signals. Thus, it is possible to detect all the fetal ORS complexes in spite of their coincidence with the maternal ECG. An averaged fetal ECG is then extracted to improve the signal-to-noise ratio, making it possible to recognize fetal P and T waves.
引用
收藏
页码:325 / 335
页数:11
相关论文
共 17 条
[1]  
ABBOUD S, 1989, J ELECTROCARDIOL, V22, P238
[2]   SPECTRAL-ANALYSIS OF THE FETAL ELECTROCARDIOGRAM [J].
ABBOUD, S ;
SADEH, D .
COMPUTERS IN BIOLOGY AND MEDICINE, 1989, 19 (06) :409-415
[3]   QUANTIFICATION OF THE FETAL ELECTROCARDIOGRAM USING AVERAGING TECHNIQUE [J].
ABBOUD, S ;
BARKAI, G ;
MASHIACH, S ;
SADEH, D .
COMPUTERS IN BIOLOGY AND MEDICINE, 1990, 20 (03) :147-155
[4]   POWER SPECTRUM ANALYSIS OF FETAL HEART-RATE-VARIABILITY USING THE ABDOMINAL MATERNAL ELECTROCARDIOGRAM [J].
ABBOUD, S ;
SADEH, D .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (02) :161-164
[5]   ABDOMINAL-LEAD FETAL ELECTROCARDIOGRAPHIC R-WAVE ENHANCEMENT FOR HEART-RATE DETERMINATION [J].
AZEVEDO, S ;
LONGINI, RL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (05) :255-260
[6]   A NEW TECHNIQUE FOR THE SUPPRESSION OF THE MECG [J].
BERGVELD, P ;
MEIJER, WJH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (04) :348-354
[7]   REAL-TIME FETAL ECG RECORDING [J].
BERGVELD, P ;
KOLLING, AJ ;
PEUSCHER, JHJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (05) :505-509
[8]   EVALUATION OF AUTOCORRELATION TECHNIQUES FOR DETECTION OF FETAL ELECTROCARDIOGRAM [J].
FAVRET, AG ;
CAPUTO, AF .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1966, BM13 (01) :37-+
[9]   FETAL ELECTROCARDIOGRAM ENHANCEMENT BY TIME-SEQUENCED ADAPTIVE FILTERING [J].
FERRARA, ER ;
WIDROW, B .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1982, 29 (06) :458-460
[10]  
GENSER G, 1968, J OBSTET GYN BR COMM, V75, P941