A scintillating fiber beam-energy monitor for electron beam therapy

被引:5
|
作者
Aoyama, T
Maekoshi, H
Tsuzaka, M
Koyama, S
机构
[1] Nagoya University College of Medical Technology, Nagoya, 461, Daikominami, Higashi-ku
关键词
electron beam; beam energy; energy constancy; quality assurance; scintillating fiber;
D O I
10.1118/1.597499
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new type of beam-energy monitor is described, which has been developed to check the energies of electron beams from radiotherapy accelerators. The monitor consists of a pair of scintillating fibers and photodiode read-out circuits that measure the energy dependence of electron transmission through a wedge-shaped absorber. A linear energy response and 1% accuracy for energy constancy measurements are attained with the monitor. The monitor having advantages of simple mechanical and electronic constructions, small size, and low cost is suitable for practical use as a portable device.
引用
收藏
页码:2101 / 2102
页数:2
相关论文
共 50 条
  • [21] SIMPLE BEAM-ENERGY MEASURER FOR HEAVY-IONS
    OLSEN, DK
    MERRILL, NH
    BIAGI, SF
    PHILLIPS, WR
    BARNETT, AR
    NUCLEAR INSTRUMENTS & METHODS, 1974, 114 (03): : 615 - 617
  • [22] Characterization of One-Dimensional Fiber-Optic Scintillating Detectors for Electron-Beam Therapy Dosimetry
    Lee, Bongsoo
    Jang, Kyoung Won
    Cho, Dong Hyun
    Yoo, Wook Jae
    Shin, Sang Hun
    Tack, Gye-Rae
    Chung, Soon-Cheol
    Kim, Sin
    Cho, Hyosung
    Park, Byung Gi
    Moon, Joo Hyun
    Kim, Siyong
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (05) : 2627 - 2631
  • [23] Beam-energy and system-size dependence of the CME
    Toneev V.D.
    Voronyuk V.
    Physics of Particles and Nuclei Letters, 2011, 8 (9) : 938 - 943
  • [24] Stability of electron-beam energy monitor for quality assurance of the electron-beam energy from radiotherapy accelerators
    Chida, K
    Saito, H
    Takai, Y
    Zuguchi, M
    Mitsuya, M
    Sakakida, H
    Kohzuki, M
    Yamada, S
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 2002, 198 (03): : 197 - 201
  • [25] Detecting electron beam energy shifts with a commercially available energy monitor
    Evans, MDC
    Moftah, BA
    Olivares, M
    Podgorsak, EB
    MEDICAL PHYSICS, 2000, 27 (07) : 1661 - 1663
  • [26] Design of an innovative beam monitor for particle therapy for the simultaneous measurement of beam fluence and energy
    Sacchi, R.
    Cartiglia, N.
    Cenna, F.
    Fanola, L.
    Ferrero, M.
    Giordanengo, S.
    Marchetto, F.
    Monaco, V.
    Vignati, A.
    Anvar, M. Varasteh
    Cirio, R.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S92 - S93
  • [27] Design of An Innovative Beam Monitor for Particle Therapy for the Simultaneous Measurement of Beam Fluence and Energy
    Sacchi, R.
    Cartiglia, N.
    Cenna, F.
    Guarachi, L. Fanola
    Ferrero, M.
    Giordanengo, S.
    Marchetto, F.
    Monaco, V.
    Vignati, A.
    Anvar, M. Varasteh
    Cirio, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3581 - 3582
  • [28] Study of fast energy electron beam profile monitor system
    Hus, I
    Yu, CI
    Chu, CC
    PROCEEDINGS OF THE 1995 PARTICLE ACCELERATOR CONFERENCE, VOLS 1-5, 1996, : 2515 - 2517
  • [29] Scintillating Fiber Array to Determine the Proton Beam Range
    Hwang, U.
    Shin, D.
    Song, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [30] Beam-Energy and Centrality Dependence of Directed Flow of Identified Particles
    Shanmuganathan, Prashanth
    NUCLEAR PHYSICS A, 2016, 956 : 260 - 263