FLEXIBLE (KAPTON-BASED) MICROSENSOR ARRAYS OF HIGH-STABILITY FOR CARDIOVASCULAR APPLICATIONS

被引:91
作者
LINDNER, E
COSOFRET, VV
UFER, S
BUCK, RP
KUSY, RP
ASH, RB
NAGLE, HT
机构
[1] UNIV N CAROLINA CHAPEL HILL, DEPT CHEM, CHAPEL HILL, NC 27599 USA
[2] UNIV N CAROLINA CHAPEL HILL, DEPT ORTHODONT & BIOMED ENGN, CHAPEL HILL, NC 27599 USA
[3] N CAROLINA STATE UNIV, DEPT ELECT & COMP ENGN, RALEIGH, NC 27695 USA
来源
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS | 1993年 / 89卷 / 02期
关键词
D O I
10.1039/ft9938900361
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The design, fabrication and performance characteristics of Kapton-based planar mini and semimicro potentiometric sensors with an Ag/AgCl or a quinhydrone-based redox internal reference electrode are described. The ion-selective membranes cast from conventional and various modified PVC matrices and containing different pH-sensitive ionophores are ranked on the basis of their performances in hostile environments. The adhesive bonding strength of the different PVC membranes to the polyimide-coated Kapton substrate was quantitatively evaluated as a function of fabrication procedure and sample solution contact. The long-term stability of the electrodes was characterized by the alterations of the analytical parameters of the sensors over a period of time, as well as by determining the resistance changes of their sensing membranes.
引用
收藏
页码:361 / 367
页数:7
相关论文
共 29 条
[11]  
Ives D. J., 1961, REFERENCE ELECTRODES
[12]  
Janata J., 1989, PRINCIPLES CHEM SENS
[13]   FABRICATION, EVALUATION, AND USE OF EXTRACELLULAR K+ AND H+ ION-SELECTIVE ELECTRODES [J].
JOHNSON, TA ;
ENGLE, CL ;
KUSY, RP ;
KNISLEY, SB ;
GRAEBNER, CA ;
GETTES, LS .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 258 (04) :H1224-H1231
[14]   THIN-FILM ION-SELECTIVE SENSORS BASED ON NEUTRAL CARRIER MEMBRANES [J].
KEPLINGER, F ;
GLATZ, R ;
JACHIMOWICZ, A ;
URBAN, G ;
KOHL, F ;
OLCAYTUG, F ;
PROHASKA, OJ .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) :272-274
[15]  
KO WH, 1985, IMPLANTABLE SENSORS, P259
[16]  
KUSY RP, UNPUB
[17]  
LINDNER E, UNPUB ANAL CHEM
[18]  
MA SC, 1990, MIKROCHIM ACTA, V1, P197
[19]  
OEHME F, 1986, GIT FACHZ LAB, P595
[20]   LIFE TIME OF NEUTRAL CARRIER BASED ION-SELECTIVE LIQUID-MEMBRANE ELECTRODES [J].
OESCH, U ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1980, 52 (04) :692-700