DETERMINATION OF PROTEOLYTIC-ENZYMES BY FLOW-INJECTION ANALYSIS

被引:1
|
作者
NICOLAS, P
LAMY, A
REYMOND, S
机构
[1] Nestlé Research Centre, Nestec Ltd, CH-1000 Lausanne 26, Vers-chez-les-Blanc
关键词
D O I
10.1016/0003-2697(91)90185-V
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Quantitation of proteolytic enzymes using N-succinyl-l-Ala-l-Pro-l-Phe-p-nitroanilide has been adapted to flow-injection analysis. This procedure has been developed using two different proteases: subtilisin and chymotrypsin. For both enzymes the influence of substrate concentration on spectrophotometric response has been studied. The assay is based on the merging zones technique combined with a washing step. Results are obtained in less than 15 s and samples may be run at a rate of 90/h with good reproducibility. A linear relation between peak heights and enzyme concentrations was observed for 0-0.15 Anson unit/liter of subtilisin and for 0-30 mg/liter of a commercial preparation of chymotrypsin. The method requires only small sample volumes, and the consumption of the chromogenic substrate is reduced to a minimum by using intermittent pumping. © 1991.
引用
收藏
页码:70 / 73
页数:4