EVALUATION OF NONINVASIVE BLOOD-PRESSURE MONITORING DEVICES SPACELAB-90202 AND SPACELAB-90207 VERSUS RESTING AND AMBULATORY 24-HOUR INTRAARTERIAL BLOOD-PRESSURE

被引:147
作者
GROPPELLI, A
OMBONI, S
PARATI, G
MANCIA, G
机构
[1] UNIV MILAN,OSPED SAN GERARDO MONTE & TERAPIA MED,CATTEDRA MED INTERNA,I-20122 MILAN,ITALY
[2] OSPED MAGGIDRE,CTR FISIOL CLIN & IPERTENS,MILAN,ITALY
[3] CTR AUXOL ITALIANO PIANCAVALLO,MILAN,ITALY
关键词
BLOOD PRESSURE MONITORS; BLOOD PRESSURE MEASUREMENT; 24-HOUR BLOOD PRESSURE PROFILES; AMBULATORY BLOOD PRESSURE MONITORING;
D O I
10.1161/01.HYP.20.2.227
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial diastolic blood pressure (Spacelabs-intra-arterial differences, -0.8+/-9.2, NS, and 9.1+/-8.8 mm Hg, p<0.01, for systolic and diastolic blood pressures, respectively). In ambulatory conditions Spacelabs-intra-arterial average differences in 24-hour values were +0.4+/-5.1 mm Hg for systolic blood pressure (NS) and +14.0+/-2.9 mm Hg for diastolic blood pressure (p<0.01) when group data were considered. The performance of both Spacelabs devices was worse when assessed in individual subjects or for each hourly interval. In spite of these differences between noninvasive and intra-arterial absolute blood pressure values, however, Spacelabs 90202 and 90207 monitors were able to faithfully reflect directional hour-to-hour changes in intra-arterial blood pressure (chi-2=18.2 and chi-2=23.1 for systolic and diastolic blood pressures, respectively, p<0.01). No differences were found between the performance of the two Spacelabs devices. Thus, although the absolute accuracy of blood pressure values provided by these monitors in ambulatory subjects is still limited, they seem to be suitable for studies aimed at assessing 24-hour blood pressure profiles quantitatively as well as qualitatively.
引用
收藏
页码:227 / 232
页数:6
相关论文
共 50 条
  • [31] AMBULATORY BLOOD-PRESSURE IN PREGNANCY - COMPARISON OF THE SPACELABS-90207 AND ACCUTRACKER-II MONITORS WITH INTRAARTERIAL RECORDINGS
    BROWN, MA
    BUDDLE, ML
    BENNETT, M
    SMITH, B
    MORRIS, R
    WHITWORTH, JA
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1995, 173 (01) : 218 - 223
  • [32] Evaluation of the antihypertensive 24-hour effect of fosinopril by means of ambulatory blood pressure monitoring
    Schneider, R
    Pongratz, G
    Machnig, T
    BrandtPohlmann, M
    Bachmann, K
    NIEREN-UND HOCHDRUCKKRANKHEITEN, 1996, 25 (08) : 382 - 384
  • [33] 24-HOUR MONITORING OF BLOOD-PRESSURE AND HEART-RATE IN HEART-TRANSPLANT PATIENTS
    GIORGI, DMA
    BORTOLOTTO, LA
    SEFERIAN, P
    BOCCHI, EA
    BERNARDESSILVA, H
    PEREIRABARRETTO, AC
    BELLOTTI, G
    PILEGGI, F
    JATENE, AD
    JOURNAL OF HYPERTENSION, 1991, 9 : S340 - S341
  • [34] COMPARATIVE-EVALUATION OF THE ANTIHYPERTENSIVE EFFICACY OF ONCE-DAILY AMLODIPINE VERSUS NITRENDIPINE WITH 24-HOUR AMBULATORY BLOOD-PRESSURE MONITORING IN ESSENTIAL-HYPERTENSION
    COCA, A
    PICADO, MJ
    DELASIERRA, A
    AGUILERA, MT
    SANCHEZ, M
    LLUCH, MM
    URBANOMARQUEZ, A
    JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1993, 22 (04) : 513 - 518
  • [35] EARLY 24-HOUR BLOOD-PRESSURE ELEVATION IN NORMOTENSIVE SUBJECTS WITH PARENTAL HYPERTENSION
    RAVOGLI, A
    TRAZZI, S
    VILLANI, A
    MUTTI, E
    CUSPIDI, C
    SAMPIERI, L
    DEAMBROGGI, L
    PARATI, G
    ZANCHETTI, A
    MANCIA, G
    HYPERTENSION, 1990, 16 (05) : 491 - 497
  • [36] DOXAZOSIN AND 48-HOUR AMBULATORY BLOOD-PRESSURE MONITORING IN NORMOTENSIVE SUBJECTS
    TAKATA, Y
    YOSHIZUMI, T
    ITO, Y
    HIROTA, Y
    FUJISHIMA, M
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1994, 58 (06): : 403 - 408
  • [37] EFFECTS OF SLOW-RELEASE VERAPAMIL AND NITRENDIPINE ON OFFICE AND 24-HOUR AMBULATORY BLOOD-PRESSURE IN HYPERTENSIVE PATIENTS
    PANNARALE, G
    PUDDU, PE
    CAMPBELL, SV
    COLLAUTO, F
    STAZI, F
    REALE, A
    JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1992, 19 : S53 - S56
  • [38] CLINICAL-SIGNIFICANCE OF AMBULATORY BLOOD-PRESSURE MONITORING - EVALUATION OF SEVERITY OF HYPERTENSION, EFFICACY OF TREATMENT AND EFFECTS ON NIGHTTIME BLOOD-PRESSURE
    KATAYAMA, S
    MARUNO, Y
    ITABASHI, A
    INABA, M
    OMOTO, A
    ISHII, J
    JAPANESE HEART JOURNAL, 1991, 32 (01): : 45 - 55
  • [39] Evaluation of nocturnal blood pressure by the multi-P analysis of 24-hour ambulatory monitoring
    Lusardi, P
    Vanasia, A
    Mugellini, A
    Zoppi, A
    Preti, P
    Fogari, R
    ZEITSCHRIFT FUR KARDIOLOGIE, 1996, 85 : 121 - 123
  • [40] Variability of circadian blood pressure profile during 24-hour ambulatory blood pressure monitoring in hypertensive patients
    Salwa, Pawel
    Gorczyca-Michta, Iwona
    Kluk, Maciej
    Dziubek, Katarzyna
    Wozakowska-Kaplon, Beata
    KARDIOLOGIA POLSKA, 2014, 72 (05) : 432 - 437