EVALUATION OF NONINVASIVE BLOOD-PRESSURE MONITORING DEVICES SPACELAB-90202 AND SPACELAB-90207 VERSUS RESTING AND AMBULATORY 24-HOUR INTRAARTERIAL BLOOD-PRESSURE

被引:147
|
作者
GROPPELLI, A
OMBONI, S
PARATI, G
MANCIA, G
机构
[1] UNIV MILAN,OSPED SAN GERARDO MONTE & TERAPIA MED,CATTEDRA MED INTERNA,I-20122 MILAN,ITALY
[2] OSPED MAGGIDRE,CTR FISIOL CLIN & IPERTENS,MILAN,ITALY
[3] CTR AUXOL ITALIANO PIANCAVALLO,MILAN,ITALY
关键词
BLOOD PRESSURE MONITORS; BLOOD PRESSURE MEASUREMENT; 24-HOUR BLOOD PRESSURE PROFILES; AMBULATORY BLOOD PRESSURE MONITORING;
D O I
10.1161/01.HYP.20.2.227
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial diastolic blood pressure (Spacelabs-intra-arterial differences, -0.8+/-9.2, NS, and 9.1+/-8.8 mm Hg, p<0.01, for systolic and diastolic blood pressures, respectively). In ambulatory conditions Spacelabs-intra-arterial average differences in 24-hour values were +0.4+/-5.1 mm Hg for systolic blood pressure (NS) and +14.0+/-2.9 mm Hg for diastolic blood pressure (p<0.01) when group data were considered. The performance of both Spacelabs devices was worse when assessed in individual subjects or for each hourly interval. In spite of these differences between noninvasive and intra-arterial absolute blood pressure values, however, Spacelabs 90202 and 90207 monitors were able to faithfully reflect directional hour-to-hour changes in intra-arterial blood pressure (chi-2=18.2 and chi-2=23.1 for systolic and diastolic blood pressures, respectively, p<0.01). No differences were found between the performance of the two Spacelabs devices. Thus, although the absolute accuracy of blood pressure values provided by these monitors in ambulatory subjects is still limited, they seem to be suitable for studies aimed at assessing 24-hour blood pressure profiles quantitatively as well as qualitatively.
引用
收藏
页码:227 / 232
页数:6
相关论文
共 50 条
  • [1] VALIDATION OF THE SPACELAB-90202 AND SPACELAB-90207 DEVICES FOR AMBULATORY BLOOD-PRESSURE MONITORING BY COMPARISON WITH INTRAARTERIAL RESTING AND AMBULATORY MEASUREMENTS
    GROPPELLI, A
    OMBONI, S
    RAVOGLI, A
    VILLANI, A
    PARATI, G
    MANCIA, G
    JOURNAL OF HYPERTENSION, 1991, 9 : S334 - S335
  • [2] EFFECT OF NISOLDIPINE ON AMBULATORY BLOOD-PRESSURE UNDER 24-HOUR NONINVASIVE MONITORING
    BLAU, A
    HERZOG, D
    SHECHTER, P
    ELIAHOU, HE
    ISRAEL JOURNAL OF MEDICAL SCIENCES, 1992, 28 (10): : 688 - 693
  • [3] AMBULATORY 24-HOUR BLOOD-PRESSURE VERSUS SELF-MEASURED BLOOD-PRESSURE IN PHARMACOLOGICAL TRIALS
    MENGDEN, T
    WEISSER, B
    VETTER, W
    JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1994, 24 : S20 - S25
  • [4] AMBULATORY 24-HOUR BLOOD-PRESSURE MONITORING - SIGNIFICANCE OF BLOOD-PRESSURE VARIABILITY FOR THE PROGRESSION OF CARDIAC-HYPERTROPHY
    MACHNIG, T
    ENGELS, G
    HENNEKE, KH
    GELLERT, J
    BACHMANN, K
    ZEITSCHRIFT FUR KARDIOLOGIE, 1993, 82 (03): : 187 - 191
  • [5] AMBULATORY 24-HOUR BLOOD-PRESSURE MONITORING IN HEALTHY-CHILDREN AND ADOLESCENTS
    KRULL, F
    EHRICH, JHH
    OFFNER, G
    FROEDE, K
    BUCK, T
    MONATSSCHRIFT KINDERHEILKUNDE, 1994, 142 (02) : 110 - 117
  • [6] AMBULATORY 24-HOUR BLOOD-PRESSURE MEASUREMENTS IN PHARMACOLOGICAL STUDIES
    WEISSER, B
    MENGDEN, T
    VETTER, W
    JOURNAL OF HYPERTENSION, 1990, 8 : S87 - S92
  • [7] ELEVATED BLOOD-PRESSURE DURING THE FIRST 2 HOURS OF AMBULATORY BLOOD-PRESSURE MONITORING - A STUDY COMPARING CONSECUTIVE 24-HOUR MONITORING PERIODS
    PRASAD, N
    MACFADYEN, RJ
    OGSTON, SA
    MACDONALD, TM
    JOURNAL OF HYPERTENSION, 1995, 13 (03) : 291 - 295
  • [8] 24-HOUR BLOOD-PRESSURE MONITORING IN HEALTHY-CHILDREN
    KRULL, F
    BUCK, T
    OFFNER, G
    BRODEHL, J
    EUROPEAN JOURNAL OF PEDIATRICS, 1993, 152 (07) : 555 - 558
  • [9] 24-HOUR AMBULATORY BLOOD-PRESSURE MONITORING AND ANTIHYPERTENSIVE TREATMENT - FOCUS ON ACE-INHIBITORS
    RAVOGLI, A
    OMBONI, S
    DECESARIS, R
    FOGARI, R
    LATTUADA, S
    MONTEMURRO, G
    PALOMBO, C
    PORCELLATI, C
    RANIERI, G
    TETTAMANTI, F
    VERDECCHIA, P
    MARELLI, C
    ZANCHETTI, A
    MANCIA, G
    JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1994, 23 : S15 - S19
  • [10] CLINICAL-SIGNIFICANCE AND COST-EFFECTIVENESS OF 24-HOUR AMBULATORY BLOOD-PRESSURE MONITORING
    IMAI, Y
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 1995, 176 (01): : 1 - 15