PHOTOMETRICS - A NEW METHOD OF MEASURING THE CROSS-SECTIONAL AREA OF THE SUBGLOTTIS

被引:14
作者
JONES, NS [1 ]
IRVING, RM [1 ]
FORD, GR [1 ]
BALAKRISHNAN, A [1 ]
ALBERT, DM [1 ]
机构
[1] HOSP SICK CHILDREN,DEPT OTOLARYNGOL,LONDON WC1N 3JH,ENGLAND
来源
CLINICAL OTOLARYNGOLOGY | 1992年 / 17卷 / 02期
关键词
LARYNGEAL STENOSIS; TRACHEAL STENOSIS; INSTRUMENTATION; ENDOSCOPY;
D O I
10.1111/j.1365-2273.1992.tb01061.x
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
A new endoscopic method of measuring the cross-sectional area of a subglottic stenosis is presented. The method was tested against existing techniques using a model. By standardizing the distance from the objective of the endoscope to the stenotic area we were able to measure the cross-sectional area using a graduated grid held against the video screen. This technique was then compared with the standard one of estimating the diameter of a stenosis from the external diameter of the bronchoscope which can be passed through the stenosis. The methods were found to be equally accurate in measuring the lumen diameter. However, when measuring the minimum cross-sectional area, three of five observers were significantly better (P < 0.01) using the photometric method rather than the standard method (Wilcoxon's rank sum test for paired data). The photometeric technique of measuring the cross-sectional area of the subglottis is easily performed, non-traumatic, and allows a visual record to be kept.
引用
收藏
页码:141 / 145
页数:5
相关论文
共 15 条
[1]  
ABRAMSON AL, 1982, LARYNGOSCOPE, V92, P922
[2]  
AW K, 1982, LARYNGOSCOPE, V92, P100
[3]   TRACHEOSTOMY DECANNULATION PANIC IN CHILDREN - FACT OR FICTION [J].
BLACK, RJ ;
BALDWIN, DL ;
JOHNS, AN .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1984, 98 (03) :297-304
[4]   PEDIATRIC LARYNGOTRACHEAL STENOSIS [J].
COTTON, RT .
JOURNAL OF PEDIATRIC SURGERY, 1984, 19 (06) :699-704
[5]  
DIAMANT H, 1962, ACTA OTOLARYNGOL S, V158, P39
[6]  
FORTE V, 1986, J OTOLARYNGOL, V15, P359
[7]  
KEARNS DB, 1988, CLIN OTOLARYNGOL, V15, P53
[8]  
MacLachlan R F, 1969, J Laryngol Otol, V83, P991, DOI 10.1017/S0022215100071243
[9]   TIDAL FLOW MEASUREMENT IN THE DECISION TO DECANNULATE THE PEDIATRIC-PATIENT [J].
MALLORY, GB ;
MUTICH, R ;
REILLY, JS ;
KENNA, MA ;
MOTOYAMA, EK ;
STOOL, SE .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1985, 94 (05) :454-457
[10]  
MILLER RD, 1969, MAYO CLIN PROC, V44, P145