EVALUATION OF MENISCAL TEARS - FAST SPIN-ECHO VERSUS CONVENTIONAL SPIN-ECHO MAGNETIC-RESONANCE-IMAGING

被引:23
|
作者
ANDERSON, MW
RAGHAVAN, N
SEIDENWURM, DJ
GREENSPAN, A
DRAKE, C
机构
[1] RADIOL ASSOCIATES SACRAMENTO,SACRAMENTO,CA
[2] UNIV CALIF DAVIS,SCH MED,DEPT STAT,SACRAMENTO,CA
关键词
MAGNETIC RESONANCE IMAGING; KNEE; MENISCI; FAST SPIN ECHO;
D O I
10.1016/S1076-6332(05)80166-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. We compared the performance of fast spin-echo (FSE) with conventional spin-echo (CSE) magnetic resonance (MR) imaging sequences in the detection of meniscal tears. Methods. Seventy-three patients underwent MR examination of the knee for suspected internal derangement. Each patient was scanned with a CSE sequence and one of two FSE sequences. The primary difference between the two FSE sequences consisted of the echo train length (ETL). Thirty-seven patients (group 1) were scanned with the FSE I sequence (ETL = 8), and 36 patients (group 2) were scanned with the FSE II sequence (ETL = 4). Menisci were graded as torn or not torn on the basis of their MR appearance, The sequences were compared with each other and with the surgical findings in 31 patients who underwent arthroscopy. Results. In group 1, there was a significant discrepancy between the CSE and FSE I techniques (p = .006). The FSE I sequence detected only 11 of 19 surgically proven torn menisci as opposed to 18 of 19 detected with the CSE sequence. The FSE II sequence performed significantly better in group 2, with an accuracy equal to that of the CSE sequence. Conclusion. FSE sequences are extremely technique dependent with regard to detecting meniscal tears and should not replace CSE sequences in this setting until further studies are performed to optimize the technique.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 50 条
  • [41] A prospective comparison of brain contrast characteristics and lesion detection using single-shot fast spin-echo and fast spin-echo
    Mittal, TK
    Halpin, SFS
    Bourne, MW
    Hourihan, MD
    Perkins, T
    Sun, Y
    Tan, S
    NEURORADIOLOGY, 1999, 41 (07) : 480 - 486
  • [42] FUNCTIONAL NMR IMAGING USING FAST SPIN-ECHO AT 1.5 T
    CONSTABLE, RT
    KENNAN, RP
    PUCE, A
    MCCARTHY, G
    GORE, JC
    MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) : 686 - 690
  • [43] A COMPARISON BETWEEN FAST AND CONVENTIONAL SPIN-ECHO IN THE DETECTION OF MULTIPLE-SCLEROSIS LESIONS
    THORPE, JW
    HALPIN, SF
    MACMANUS, DG
    BARKER, GJ
    KENDALL, BE
    MILLER, DH
    NEURORADIOLOGY, 1994, 36 (05) : 388 - 392
  • [44] T2-weighted MRI of rectosigmoid carcinoma: Comparison of respiratory-triggered fast spin-echo, breathhold fast-recovery fast spin-echo, and breathhold single-shot fast spin-echo sequences
    Yamashita, Shuhei
    Masui, Takayuki
    Katayama, Motoyuki
    Sato, Kimihiko
    Yoshizawa, Nobuko
    Seo, Hidekazu
    Sakahara, Harumi
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 511 - 516
  • [45] DIAGNOSIS OF PARTIAL AND COMPLETE ROTATOR CUFF TEARS USING COMBINED GRADIENT-ECHO AND SPIN-ECHO IMAGING
    TUITE, MJ
    YANDOW, DR
    DESMET, AA
    ORWIN, JF
    QUINTANA, FA
    SKELETAL RADIOLOGY, 1994, 23 (07) : 541 - 545
  • [46] Non-invasive temperature imaging of muscles with magnetic resonance imaging using spin-echo sequences
    E. Mietzsch
    M. Koch
    M. Schaldach
    J. Werner
    B. Bellenberg
    K. U. Wentz
    Medical and Biological Engineering and Computing, 1998, 36 : 673 - 678
  • [47] Non-invasive temperature imaging of muscles with magnetic resonance imaging using spin-echo sequences
    Mietzsch, E
    Koch, M
    Schaldach, M
    Werner, J
    Bellenberg, B
    Wentz, KU
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (06) : 673 - 678
  • [48] Fast spin echo vs conventional spin echo in cervical spine imaging
    Gillams, AR
    Soto, JA
    Carter, AP
    EUROPEAN RADIOLOGY, 1997, 7 (08) : 1211 - 1214
  • [49] SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING
    BAKKER, CJG
    BHAGWANDIEN, R
    MOERLAND, MA
    RAMOS, LMP
    MAGNETIC RESONANCE IMAGING, 1994, 12 (05) : 767 - 774
  • [50] HIGH-RESOLUTION MAGNETIC-RESONANCE-IMAGING OF THE ANTERIOR VISUAL PATHWAY IN PATIENTS WITH OPTIC NEUROPATHIES USING FAST SPIN-ECHO AND PHASED-ARRAY LOCAL COILS
    GASS, A
    BARKER, GJ
    MACMANUS, D
    SANDERS, M
    RIORDANEVA, P
    TOFTS, PS
    THORPE, J
    MCDONALD, WI
    MOSELEY, IF
    MILLER, DH
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1995, 58 (05) : 562 - 569