Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX

被引:18
作者
Manikandan, A. [1 ]
Biplab, Sarkar [2 ]
Perianayagam, David A. [1 ]
Holla, R. [1 ]
Vivek, T. R. [3 ]
Sujatha, N. [4 ]
机构
[1] Narayana Hrudayalaya, Dept Radiat Oncol, Bangalore, Karnataka, India
[2] AMRI Hosp, Kolkata, India
[3] Jupiter Hosp, Mumbai, Maharashtra, India
[4] GGH, Dept Radiotherapy, Guntur, India
关键词
2D array; brachy; IMatriXX;
D O I
10.4103/0971-6203.83491
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of -0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error <= 1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size <= 1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile.
引用
收藏
页码:171 / 175
页数:5
相关论文
共 10 条
[1]   AN INTEGRATED PHANTOM FOR HDR QUALITY-CONTROL [J].
ALDRICH, JE ;
SAMANT, SS .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (784) :363-365
[2]   Dosimetric characterization of a large area pixel-segmented ionization chamber [J].
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Fidanzio, A ;
Garelli, E ;
Giordanengo, S ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Piermattei, A ;
Freire, CJS ;
Sardo, A ;
Trevisiol, E .
MEDICAL PHYSICS, 2004, 31 (02) :414-420
[3]   Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine [J].
Herzen, J. ;
Todorovic, M. ;
Cremers, F. ;
Platz, V. ;
Albers, D. ;
Bartels, A. ;
Schmidt, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (04) :1197-1208
[4]  
Jonathan G, 2009, J APPL CLIN MED PHYS, V10, P62
[5]   A quality assurance test tool for high dose-rate remote afterloading brachytherapy units [J].
Li, ZF ;
Mitchell, TP ;
Palta, JR ;
Liu, CR .
MEDICAL PHYSICS, 1998, 25 (02) :232-235
[6]   Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56 [J].
Nath, R ;
Anderson, LL ;
Meli, JA ;
Olch, AJ ;
Stitt, JA ;
Williamson, JF .
MEDICAL PHYSICS, 1997, 24 (10) :1557-1598
[7]   A quality assurance tool for high-dose-rate brachytherapy [J].
Rickey, Daniel W. ;
Sasaki, David ;
Bews, Jeff .
MEDICAL PHYSICS, 2010, 37 (06) :2525-2532
[8]   Measurement of transit time of a remote after-loading high dose rate brachytherapy source [J].
Sahoo, N .
MEDICAL PHYSICS, 2001, 28 (08) :1786-1790
[9]   Tracking brachytherapy sources using emission imaging with one flat panel detector [J].
Song, Haijun ;
Bowsher, James ;
Das, Shiva ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2009, 36 (04) :1109-1111
[10]   Dosimetric verification of brain and head and neck intensity-modulated radiation therapy treatment using EDR2 films and 2D ion chamber array matrix [J].
Varatharaj, C. ;
Ravikumar, M. ;
Sathiyan, S. ;
Supe, Sanjay S. ;
Vivek, T. R. ;
Manikandan, A. .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2010, 6 (02) :179-184