MRI SCANNER TRANSPORT-SYSTEM

被引:0
作者
PHILLIPS, JR [1 ]
FALCONER, JC [1 ]
BOTZER, S [1 ]
NARAYANA, PA [1 ]
机构
[1] UNIV TEXAS,SCH MED,DEPT RADIOL,HOUSTON,TX 77030
关键词
TRANSPORT SYSTEM; MAGNETIC RESONANCE SCANNER;
D O I
10.1002/mrm.1910320119
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A computer-controlled transport system has been designed and tested for small bore magnetic resonance scanners for animal research. The transport system allows remote animal positioning based on scout images using a graphic interface. The design utilized commercially available components suitable for use in the scanner's strong magnetic field. Tests indicate positioning accuracy of the transport system to be within 0.13 mm.
引用
收藏
页码:133 / 136
页数:4
相关论文
共 2 条
  • [1] INVIVO P-31 NMR-STUDIES OF AVIAN DYSTROPHIC MUSCLES
    NARAYANA, PA
    DELAYRE, JL
    MISRA, LK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) : 549 - 553
  • [2] 1990, INSTALLATION USER RE