PLAQUE ABRASION AND INTRAAORTIC BALLOON LEAK

被引:8
作者
COX, PM [1 ]
GORAN, SF [1 ]
MORTON, J [1 ]
WOLVEK, S [1 ]
机构
[1] DATASCOPE CORP,FAIRFIELD,NJ
关键词
COMPLICATIONS; INTRAAORTIC BALLOON; RUPTURE; LEAK;
D O I
10.1378/chest.108.6.1495
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objective: To determine and correct cause of high incidence of intra-aortic balloon leaks (ruptures). Design: Epidemiologic investigation of factors associated with intra-aortic balloon leak, and sequential application of corrective measures evaluated by continued concurrent data collection. Setting: Thirty-four-bed ICU in 598-bed tertiary care medical center. Intercentions: Procedure changed to place smaller balloons (34 mL instead of 40 mL) in patients less than 163 cm in height. Measurements and results: Demographic and clinical data on all patients showed no change after initial interventions, followed by significant drop (8 to 2%) in aorta. incidence of balloon leak when smaller, shorter balloons were placed in shorter patients. Conclusions: Placement of larger, longer balloons in patients increases risk of perforation of balloon by calcific plaque in the distal thoracic and abdominal aorta.
引用
收藏
页码:1495 / 1498
页数:4
相关论文
共 7 条
[1]  
CURTIS JJ, 1988, AM SURGEON, V54, P142
[2]   OPERATIVE INTRA-AORTIC BALLOON RUPTURE [J].
FINEGAN, BA ;
COMM, DG .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1988, 35 (03) :297-299
[3]   ARTERIAL HELIUM EMBOLISM FROM A RUPTURED INTRAAORTIC BALLOON [J].
FREDERIKSEN, JW ;
SMITH, J ;
BROWN, P ;
ZINETTI, C .
ANNALS OF THORACIC SURGERY, 1988, 46 (06) :690-692
[4]   COMPLICATIONS OF THE INTRA-AORTIC BALLOON COUNTERPULSATION DEVICE - CLINICAL AND MORPHOLOGIC OBSERVATIONS IN 45 NECROPSY PATIENTS [J].
ISNER, JM ;
COHEN, SR ;
VIRMANI, R ;
LAWRINSON, W ;
ROBERTS, WC .
AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02) :260-268
[5]   INTRAAORTIC BALLOON PUMPING 1967 THROUGH 1982 - ANALYSIS OF COMPLICATIONS IN 733 PATIENTS [J].
KANTROWITZ, A ;
WASFIE, T ;
FREED, PS ;
RUBENFIRE, M ;
WAJSZCZUK, W ;
SCHORK, MA .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (11) :976-983
[6]  
KVILEKVAL KHV, 1991, ARCH SURG-CHICAGO, V126, P621
[7]   Events associated with rupture of intra-aortic balloon counterpulsation devices [J].
Sutter, F.P. ;
Joyce, D.H. ;
Bailey, B.M. ;
Laub, G.W. ;
Fernandez, J. ;
Pollock, S.B. ;
Adkins, M.S. ;
McGrath, L.B. .
ASAIO Transactions, 1991, 37 (01) :38-40