REDUCTION OF GLUCOSE INTERFERENCE IN BIOCHEMICAL ANALYSES BASED ON INFRARED-SPECTROSCOPY

被引:0
作者
FUJII, T
MIYAHARA, Y
WATANABE, Y
机构
关键词
INFRARED SPECTROSCOPY; ATTENUATED TOTAL REFLECTION; BLOOD SERUM; MULTIVARIATE CALIBRATION; PARTIAL LEAST SQUARES;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Simultaneous analyses of biochemical components in human serum were investigated using attenuated total reflection (ATR) and FT-IR in combination with the partial least squares (PLS) calibration method. When the glucose concentration in a sample was out of range of that in the standard solutions for calibration, measurements of biochemical components were found to be affected by the glucose concentration. Glucose interference could be reduced by optimizing the number of standard solutions and the range of glucose concentrations in the standard solutions. When the optimized standard solutions were used for calibration, good correlations between the ATR/FT-IR method and the automatic clinical analyzer could be obtained for measurements of triglycerides, urea, cholesterol, total protein and glucose, while correlation coefficients for uric acid and creatinine were 0.77 and 0.85 respectively because of insufficient sensitivity of the detector.
引用
收藏
页码:465 / 469
页数:5
相关论文
共 9 条
[1]   MONITORING OF GLUCOSE IN BIOLOGICAL-FLUIDS BY FOURIER-TRANSFORM INFRARED SPECTROMETRY WITH A CYLINDRICAL INTERNAL REFLECTANCE CELL [J].
BAUER, B ;
FLOYD, TA .
ANALYTICA CHIMICA ACTA, 1987, 197 :295-301
[2]   GLUCOSE DETERMINATION IN SIMULATED BLOOD-SERUM SOLUTIONS BY FOURIER-TRANSFORM INFRARED-SPECTROSCOPY - INVESTIGATION OF SPECTRAL INTERFERENCES [J].
BHANDARE, P ;
MENDELSON, Y ;
STOHR, E ;
PEURA, RA .
VIBRATIONAL SPECTROSCOPY, 1994, 6 (03) :363-378
[3]   PARTIAL LEAST-SQUARES REGRESSION - A TUTORIAL [J].
GELADI, P ;
KOWALSKI, BR .
ANALYTICA CHIMICA ACTA, 1986, 185 :1-17
[4]   MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD BY ATTENUATED TOTAL REFLECTION INFRARED-SPECTROSCOPY [J].
HEISE, HM ;
MARBACH, R ;
JANATSCH, G ;
KRUSEJARRES, JD .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2009-2015
[5]   MULTICOMPONENT ASSAY FOR BLOOD SUBSTRATES IN HUMAN PLASMA BY MIDINFRARED SPECTROSCOPY AND ITS EVALUATION FOR CLINICAL ANALYSIS [J].
HEISE, HM ;
MARBACH, R ;
KOSCHINSKY, T ;
GRIES, FA .
APPLIED SPECTROSCOPY, 1994, 48 (01) :85-95
[6]   MULTIVARIATE CALIBRATION FOR ASSAYS IN CLINICAL-CHEMISTRY USING ATTENUATED TOTAL REFLECTION INFRARED-SPECTRA OF HUMAN-BLOOD PLASMA [J].
JANATSCH, G ;
KRUSEJARRES, JD ;
MARBACH, R ;
HEISE, HM .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2016-2023
[7]   Improved selectivity in spectroscopy by multivariate calibration [J].
Martens, Harald ;
Karstang, Terje ;
Næs, Tormod .
Journal of Chemometrics, 1987, 1 (04) :201-219
[8]   POSTPRANDIAL BLOOD-GLUCOSE DETERMINATION BY QUANTITATIVE MIDINFRARED SPECTROSCOPY [J].
WARD, KJ ;
HAALAND, DM ;
ROBINSON, MR ;
EATON, RP .
APPLIED SPECTROSCOPY, 1992, 46 (06) :959-965
[9]   BLOOD-GLUCOSE MEASUREMENT BY INFRARED-SPECTROSCOPY [J].
ZELLER, H ;
NOVAK, P ;
LANDGRAF, R .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (02) :129-135