FIRST MEASUREMENTS WITH A DIAMOND MICROSTRIP DETECTOR

被引:49
|
作者
BORCHELT, F
DULINSKI, W
GAN, KK
HAN, S
HASSARD, J
HOWARD, A
KAGAN, H
KANIA, DR
KASS, R
LU, G
NYGARD, E
PAN, LS
SCHNETZER, S
STONE, R
STRAVER, J
TESAREK, RJ
TRISCHUK, W
WEILHAMMER, P
WHITE, C
WOODIN, RL
ZHAO, S
机构
[1] CERN,CH-1211 GENEVA 23,SWITZERLAND
[2] UNIV LONDON IMPERIAL COLL SCI & TECHNOL,LONDON SW7 2AZ,ENGLAND
[3] CRN,LEPSI,F-67037 STRASBOURG,FRANCE
[4] LAWRENCE LIVERMORE NATL LAB,LIVERMORE,CA 94551
[5] OHIO STATE UNIV,COLUMBUS,OH 43210
[6] UNIV OSLO,N-0316 OSLO 3,NORWAY
[7] RUTGERS STATE UNIV,PISCATAWAY,NJ 08855
[8] ST GOBAIN NORTON DIAMOND FILM,NORTHBOROUGH,MA 01532
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 1995年 / 354卷 / 2-3期
关键词
D O I
10.1016/0168-9002(94)01016-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have constructed the first high resolution strip detector using chemical vapor deposited diamond as the detection medium. Devices produced with this material have the possibility of being extremely radiation hard with direct applications at high luminosity colliders. This paper details the detector material, the low noise readout electronics and the detector module. First results from a test with high momentum charged particles in a testbeam at CERN are described. We achieved a signal-to-noise of 6:1 and an efficiency of 85% for minimum ionizing particles in the testbeam. The detector has a strip pitch of 100 mu m and a strip width of 50 mu m. The measured position resolution we achieved was sigma = 26 mu m Future development of diamond detectors with application in particle physics experiments and other fields is discussed.
引用
收藏
页码:318 / 327
页数:10
相关论文
共 50 条
  • [1] First measurements with a diamond microstrip detector
    Nucl Instrum Methods Phys Res Sect A, 2-3 (318):
  • [2] First neutron spectroscopy measurements with a pixelated diamond detector at JET
    Muraro, A.
    Giacomelli, L.
    Nocente, M.
    Rebai, M.
    Rigamonti, D.
    Belli, F.
    Calvani, P.
    Figueiredo, J.
    Girolami, M.
    Gorini, G.
    Grosso, G.
    Murari, A.
    Popovichev, S.
    Trucchi, D. M.
    Tardocchi, M.
    Abhangi, M.
    Abreu, P.
    Aftanas, M.
    Afzal, M.
    Aggarwal, K. M.
    Aho-Mantila, L.
    Ahonen, E.
    Aints, M.
    Airila, M.
    Albanese, R.
    Alegre, D.
    Alessi, E.
    Aleynikov, P.
    Alfier, A.
    Alkseev, A.
    Allan, P.
    Almaviva, S.
    Alonso, A.
    Alper, B.
    Alsworth, I.
    Alves, D.
    Ambrosino, G.
    Ambrosino, R.
    Amosov, V.
    Andersson, F.
    Andersson Sunden, E.
    Angelone, M.
    Anghel, A.
    Anghel, M.
    Angioni, C.
    Appel, L.
    Apruzzese, G.
    Arena, P.
    Ariola, M.
    Arnichand, H.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):
  • [3] Diamond microstrip detector for deep UV imaging
    Ciancaglioni, I
    Spaziani, F
    Rossi, MC
    Conte, G
    Kononenko, V
    Ralchenko, V
    DIAMOND AND RELATED MATERIALS, 2005, 14 (3-7) : 526 - 530
  • [4] First Measurements with the DTAS Detector
    Guadilla, V.
    Algora, A.
    Tain, J. L.
    BASIC CONCEPTS IN NUCLEAR PHYSICS: THEORY, EXPERIMENTS AND APPLICATIONS, 2016, 182 : 173 - 175
  • [5] Measurements on a GaAs microstrip detector with 50 GeV/c electrons
    Braunschweig, W
    Karpinski, W
    Kubicki, T
    Lubelsmeyer, K
    Ortmanns, J
    Pandoulas, D
    Pierschel, G
    Rente, C
    Siedling, R
    Syben, O
    Tenbusch, F
    Toporowsky, M
    Wilms, T
    Wittmer, B
    Xiao, WJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 370 (2-3): : 377 - 380
  • [6] Fabrication and testing of a microstrip particle detector based on highly oriented diamond films
    Han, SK
    McClure, MT
    Wolden, CA
    Vlahovic, B
    Soldi, A
    Sitar, S
    DIAMOND AND RELATED MATERIALS, 2000, 9 (3-6) : 1008 - 1012
  • [7] First measurements of triton burnup neutron spectra using a natural diamond detector on JT-60U
    Isobe, M
    Nishitani, T
    Krasilnikov, AV
    Kaneko, J
    Sasao, M
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 573 - 576
  • [8] Development of CVD diamond detector for time-of-flight measurements
    Michimasa, S.
    Takaki, M.
    Dozono, M.
    Go, S.
    Baba, H.
    Ideguchi, E.
    Kisamori, K.
    Matsubara, H.
    Miya, H.
    Ota, S.
    Sakai, H.
    Shimoura, S.
    Stolz, A.
    Tang, T. L.
    Tokieda, H.
    Uesaka, T.
    Zegers, R. G. T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 317 : 710 - 713
  • [9] Experimental neutron flux measurements with a diamond detector at the QUINTA setup
    Berlev A.I.
    Rodionov N.B.
    Tyutyunnikov S.I.
    Amosov V.N.
    Meshchaninov S.A.
    Yudin I.P.
    Physics of Particles and Nuclei Letters, 2016, 13 (3) : 352 - 357
  • [10] Leakage current measurements of a pixelated polycrystalline CVD diamond detector
    Zain, R. M.
    Maneuski, D.
    O'Shea, V.
    Bates, R.
    Blue, A.
    Cunnigham, L.
    Stehl, C.
    Berderman, E.
    Rahim, R. A.
    JOURNAL OF INSTRUMENTATION, 2013, 8