Three types of flow-through, three-electrode cells with porous working electrodes made of crushed vitreous carbon were tested. The electrochemical properties of the cells were studied by hydrodynamic and cyclic voltammetry using potassium ferrocyanide test solutions. Depending on the volume of the porous working electrode, Nernstian behaviour and electrochemical yields up to 100% were observed for flow rates 1 to 7 ml/min. In the stopped-flow regime the cells exhibit thin-layer cell properties. The utility of the cells is demonstrated for on-line coulometric detection, continuous removal of impurities and dissolved oxygen and for anodic stripping coulometry of trace copper.