2 CONFIGURATIONS OF THE 4-RING BIRDCAGE COIL FOR H-1 IMAGING AND H-1-DECOUPLED P-31 SPECTROSCOPY OF THE HUMAN HEAD

被引:68
作者
MURPHYBOESCH, J
SRINIVASAN, R
CARVAJAL, L
BROWN, TR
机构
[1] Department of NMR and Medical Spectroscopy, Fox Chase Cancer Center, Philadelphia, PA 19111
来源
JOURNAL OF MAGNETIC RESONANCE SERIES B | 1994年 / 103卷 / 02期
关键词
D O I
10.1006/jmrb.1994.1017
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The four-ring birdcage resonator, a new class of dual-tuned birdeage resonators, is described. We report two configurations of the coil: the low-pass, high-pass (LP-HP) and the low-pass, low-pass (LP-LP), both of which can be operated in dual quadrature mode at 1.5 T. As head coils, both configurations exhibit greatly reduced tuning interactions between frequencies, permitting rapid, noniterative tuning. Compared with single-tuned, two-ring birdcage resonators of similar volume, the sensitivity and transmitter efficiencies of the resonators are better than 85% for the proton frequency and the same to within 5% for the phosphorus frequency. Circuit models have been developed to refine coil tuning and aid the calculation of B1 field contour plots. Both configurations have been used for integrated examinations involving acquisition of high-quality 1H images and 1H-decoupled 31P CSI spectra of the human head. A scaled-down version of the LP-LP configuration has been demonstrated for use with the human calf. © 1994 Academic Press, Inc.
引用
收藏
页码:103 / 114
页数:12
相关论文
共 22 条
[1]   EFFICIENT DECOUPLER COIL DESIGN WHICH REDUCES HEATING IN CONDUCTIVE SAMPLES IN SUPERCONDUCTING SPECTROMETERS [J].
ALDERMAN, DW ;
GRANT, DM .
JOURNAL OF MAGNETIC RESONANCE, 1979, 36 (03) :447-451
[2]   HUMAN INVIVO PHOSPHATE METABOLITE IMAGING WITH P-31 NMR [J].
BOTTOMLEY, PA ;
CHARLES, HC ;
ROEMER, PB ;
FLAMIG, D ;
ENGESETH, H ;
EDELSTEIN, WA ;
MUELLER, OM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (03) :319-336
[3]  
CHEN CN, 1963, J MAGN RESON, V54, P324
[4]  
COLLIN RE, 1966, F MICROWAVE ENG, P336
[5]   DESIGN AND EVALUATION OF A NOVEL DUAL-TUNED RESONATOR FOR SPECTROSCOPIC IMAGING [J].
DERBY, K ;
TROPP, J ;
HAWRYSZKO, C .
JOURNAL OF MAGNETIC RESONANCE, 1990, 86 (03) :645-651
[6]  
FITZSIMMONS JR, 1990, 9TH SOC MAGN RES MED, V1, P532
[7]   COMPARISON OF LINEAR AND CIRCULAR-POLARIZATION FOR MAGNETIC-RESONANCE IMAGING [J].
GLOVER, GH ;
HAYES, CE ;
PELC, NJ ;
EDELSTEIN, WA ;
MUELLER, OM ;
HART, HR ;
HARDY, CJ ;
ODONNELL, M ;
BARBER, WD .
JOURNAL OF MAGNETIC RESONANCE, 1985, 64 (02) :255-270
[8]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[9]   A DESIGN FOR A DOUBLE-TUNED BIRDCAGE COIL FOR USE IN AN INTEGRATED MRI/MRS EXAMINATION [J].
ISAAC, G ;
SCHNALL, MD ;
LENKINSKI, RE ;
VOGELE, K .
JOURNAL OF MAGNETIC RESONANCE, 1990, 89 (01) :41-50
[10]  
KELLEY DAC, 1989, 8TH SOC MAGN RES MED, V2, P945