The Accuracy of Fit of Crowns Made From Wax Patterns Produced Conventionally (Hand Formed) and Via CAD/CAM Technology

被引:37
作者
Fathi, Hawa M. [1 ]
Al-Masoody, Azal H. [1 ]
El-Ghezawi, Narjes [1 ]
Johnson, Anthony [1 ]
机构
[1] Univ Sheffield, Acad Unit Restorat Dent, Sch Clin Dent, Sheffield S10 2TA, S Yorkshire, England
关键词
Marginal Gap; Crowns; Internal Fit; CAD-CAM; Cementation;
D O I
10.1922/EJPRD_1444Fathi08
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: the aim of this study was to compare the marginal and internal fit accuracy of crowns constructed using three different wax production methods; conventional, milled, and 3D printed and utilizing three different cement gap thicknesses. Methods: 15 identical stone dies were made for each method. Wax patterns were produced and then cast into metal crowns, which were assessed for differences in accuracy of both marginal and internal fit. Measurement points were 4 points per crown; occlusal, axial, marginal gap, and marginal discrepancy points. A silicone impression technique and conventional cementation technique were employed to facilitate the measurements. Results: there was no significant difference (p>0.05) in accuracy between the digitally and manually produced crowns, with the 3D printed crowns being slightly more accurate. Whereas, statistically significant differences were noticed between the conventional versus printed groups, in occlusal and marginal gap points (p<0.05). Conclusion: The marginal and internal fit of 3D printed wax patterns is more accurate than the other two production methods. The milling of wax crowns is as accurate as the conventional hand carved production in terms of internal and marginal fit. The manufacturer recommended offset/die-spacer of 30 mu m produced the most accurate internal and marginal fits.
引用
收藏
页码:P10 / P17
页数:8
相关论文
共 28 条
  • [1] Aditya Priyam, 2012, J Indian Prosthodont Soc, V12, P143, DOI 10.1007/s13191-012-0119-8
  • [2] Barnes H. A., 1989, INTRO RHEOLOGY
  • [3] Clinical fit of Procera AllCeram crowns
    Boening, KW
    Wolf, BH
    Schmidt, AE
    Kästner, K
    Walter, MH
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2000, 84 (04) : 419 - 424
  • [4] EVALUATION OF DETAIL REPRODUCTION FOR 3 DIE MATERIALS BY USING SCANNING ELECTRON-MICROSCOPY AND 2-DIMENSIONAL PROFILOMETRY
    DERRIEN, G
    LEMENN, G
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1995, 74 (01) : 1 - 7
  • [5] Determination of the minimum number of marginal gap measurements required for practical in vitro testing
    Groten, M
    Axmann, D
    Pröbster, L
    Weber, H
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2000, 83 (01) : 40 - 49
  • [6] Hobo S, 1997, FUNDAMENTALS FIXED P
  • [7] CONSIDERATIONS IN MEASUREMENT OF MARGINAL FIT
    HOLMES, JR
    BAYNE, SC
    HOLLAND, GA
    SULIK, WD
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1989, 62 (04) : 405 - 408
  • [8] GINGIVAL MARGINS FOR CROWNS - A REVIEW AND DISCUSSION .2. DISCREPANCIES AND CONFIGURATIONS
    HUNTER, AJ
    HUNTER, AR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1990, 64 (06) : 636 - 642
  • [9] Johnson T, 2016, BASICS DENT TECHNOLO, V2nd, P87
  • [10] Clinical evaluation of the marginal fit of cast crowns - validation of the silicone replica method
    Laurent, M.
    Scheer, P.
    Dejou, J.
    Laborde, G.
    [J]. JOURNAL OF ORAL REHABILITATION, 2008, 35 (02) : 116 - 122