A RELIABLE AND ACCURATE PROCEDURE FOR PREPARING LOW-ACTIVITY EFFICIENCY CALIBRATION STANDARDS FOR GERMANIUM GAMMA-RAY SPECTROMETERS

被引:4
作者
CROUDACE, IW
机构
[1] Department of Geology, University of Southampton, Southampton
来源
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-LETTERS | 1991年 / 153卷 / 02期
关键词
D O I
10.1007/BF02164875
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An empirical procedure is described which is readily capable of producing reliable and accurate efficiency calibration sources of low activity for high-resolution gamma-ray spectrometers. Known activities of a multi-isotopic solution are adsorbed onto a mixture of cation and anion-exchange resins and chromatographic-grade cellulose. The dried powder that results can be mixed with any desired solid natural matrix material to produce a homogeneous, extended geometry, efficiency calibration source. Results are presented which enable the method to be critically evaluated.