MEASUREMENT OF K-SHELL PHOTOELECTRIC CROSS-SECTIONS USING A DIRECT METHOD

被引:3
|
作者
SASI, PR [1 ]
NARAYANA, VL [1 ]
PARTHASARDHI, K [1 ]
NARASIMHAM, KL [1 ]
REDDY, SB [1 ]
RAMANIAH, KV [1 ]
机构
[1] ANDHRA UNIV, DEPT NUCL PHYS, WALTAIR 530003, ANDHRA PRADESH, INDIA
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 1987年 / 256卷 / 02期
关键词
RADIATION DETECTORS - SPECTROMETERS - SPECTRUM ANALYSIS;
D O I
10.1016/0168-9002(87)90235-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
K-shell photoelectric cross sections are determined by the direct method at 661. 65 keV in Au at 30 degree and 40 degree and in Yb, Ta and U at 40 degree . A high resolution HPGe detector is used to detect both the X-ray and incoherently scattered gamma ray intensities simultaneously. The superiority of the direct method over the indirect method at higher energies is discussed. The results are showing a good agreement with the available theoretical and other experimental values. A close observation shows that the measured values are in a better agreement with the accurate and recently reported data of Scofield.
引用
收藏
页码:373 / 376
页数:4