BIOSENSORS FOR DETECTION OF ENZYMES IMMOBILIZED IN MICROVOLUME REACTION CHAMBERS

被引:24
作者
BOUSSE, L
KIRK, G
SIGAL, G
机构
[1] Molecular Devices Corporation, Menlo Park, CA 94025
关键词
D O I
10.1016/0925-4005(90)80272-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A biosensor system is presented in which a small quantity of enzyme placed in a very small volume produces a measurable chemical change. The enzyme used is urease, and the pH changes it causes are measured with a light-addressable potentiometric sensor (LAPS). A model is presented which prediocts the sensitivity of the system as a function of chamber volume, specific activity of the enzyme, and the amount of pH buffering present. Under our usual conditions, the detection limit for urease was 4.4 × 107 molecules, which corresponds to 40 pg or (in a 300 μl sample volume) a concentration of 2.4 × 10-13 M. Much lower detection limits have been achieved by using smaller chambers and lower buffering. The urease is immobilized on a porous membrane which is then placed into the reading chamber. By making conjugates in which urease is linked to antibodies or binding proteins to a desired analyte, an assay for any analyte can be constructed. Results on an assay for total single-stranded DNA show a detection limit of 2 pg. © 1990.
引用
收藏
页码:555 / 560
页数:6
相关论文
共 13 条
[1]  
Bates R, 1972, DETERMINATION PH THE
[2]   TIME-DEPENDENCE OF THE CHEMICAL RESPONSE OF SILICON-NITRIDE SURFACES [J].
BOUSSE, L ;
HAFEMAN, D ;
TRAN, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) :361-367
[3]  
BRIGGS J, 1989, AM BIOTECHNOL LA FEB
[4]  
Butt W. R., 1984, PRACTICAL IMMUNOASSA
[5]   FIELD-EFFECT TRANSISTOR SENSITIVE TO PENICILLIN [J].
CARAS, S ;
JANATA, J .
ANALYTICAL CHEMISTRY, 1980, 52 (12) :1935-1937
[6]   PH-BASED ENZYME POTENTIOMETRIC SENSORS .3. PENICILLIN-SENSITIVE FIELD-EFFECT TRANSISTOR [J].
CARAS, SD ;
JANATA, J .
ANALYTICAL CHEMISTRY, 1985, 57 (09) :1924-1925
[7]   PH-BASED ENZYME POTENTIOMETRIC SENSORS .2. GLUCOSE-SENSITIVE FIELD-EFFECT TRANSISTOR [J].
CARAS, SD ;
PETELENZ, D ;
JANATA, J .
ANALYTICAL CHEMISTRY, 1985, 57 (09) :1920-1923
[8]   PH-BASED ENZYME POTENTIOMETRIC SENSORS .1. THEORY [J].
CARAS, SD ;
JANATA, J ;
SAUPE, D ;
SCHMITT, K .
ANALYTICAL CHEMISTRY, 1985, 57 (09) :1917-1920
[9]   A CRITICAL-EVALUATION OF THE MECHANISM OF POTENTIAL RESPONSE OF ANTIGEN POLYMER MEMBRANES TO THE CORRESPONDING ANTISERUM [J].
COLLINS, S ;
JANATA, J .
ANALYTICA CHIMICA ACTA, 1982, 136 (APR) :93-99
[10]   LIGHT-ADDRESSABLE POTENTIOMETRIC SENSOR FOR BIOCHEMICAL SYSTEMS [J].
HAFEMAN, DG ;
PARCE, JW ;
MCCONNELL, HM .