Agreement of noninvasive tear break-up time measurement between Tomey RT-7000 Auto Refractor-Keratometer and Oculus Keratograph 5M

被引:35
作者
Lee, Ryan [1 ,2 ]
Yeo, Sharon [1 ]
Aung, Han Tun [3 ]
Tong, Louis [1 ,2 ,4 ,5 ]
机构
[1] Natl Univ Singapore, Ocular Surface Res Grp, Singapore Eye Res Inst, Singapore, Singapore
[2] Natl Univ Singapore, Yong Loo Lin Sch Med, Singapore, Singapore
[3] Ngee Ann Polytech, Sch Hlth Sci, Singapore, Singapore
[4] Duke NUS Med Sch, Singapore, Singapore
[5] Singapore Natl Eye Ctr, Cornea & External Eye Dis Serv, 11 Third Hosp Ave, Singapore 168751, Singapore
来源
CLINICAL OPHTHALMOLOGY | 2016年 / 10卷
关键词
tear break-up time; tear film; Oculus Keratograph; Tomey RT-7000; dry eye;
D O I
10.2147/OPTH.S110180
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: It is difficult to standardize assessment of dry eye in different clinical settings. Increasingly, tear stability is recognized to be important for the definition and assessment of patients with dry eye. Recently, two commercially available instruments have been made available for objectively measuring noninvasive tear break-up time (NIBUT), as an indicator of tear stability: the Tomey RT-7000 Auto Refractor-Keratometer and Oculus Keratograph (K) 5M. We aim to assess the agreement of NIBUT measurements using these modalities. Methods: This prospective cross-sectional study was carried out in a tertiary referral eye center and involved 126 consecutive dry eye patients. NIBUT assessment was performed on the right eyes of participants with both the RT-7000 and the K5M techniques, with the order of assessment randomized. The Standardized Patient Evaluation of Eye Dryness (SPEED) questionnaires were administered to assess dry eye symptoms in the 2 weeks before assessment. Results: The age of the participants was 56.0 +/- 14.3 years (69.84% females). Measurements for both modalities were non-normally distributed (right-skewed). The median RT-7000 and K5M readings were 4.2 (range 0.1-10.0) and 6.4 (0.1-24.9) seconds, respectively. RT-7000 and K5M readings were poorly correlated (rho=0.061, P=0.495). Intraclass correlation coefficient between the modalities was 0.187 (95% confidence interval -0.097 to 0.406). The Bland-Altman plot showed no systematic differences between the readings with these machines. The agreement between machines was not different in different SPEED categories. Conclusion: While there are theoretical and practical benefits of NIBUT for assessment of tear stability over dye-based methods, the agreement between the two modalities was poor. Hence, studies and trials assessing NIBUT should avoid using these modalities interchangeably for NIBUT assessment. More research is needed to improve consensus on how to determine NIBUT.
引用
收藏
页码:1785 / 1790
页数:6
相关论文
共 21 条
  • [1] Noninvasive Imaging of Tear Film Dynamics in Eyes With Ocular Surface Disease
    Abdelfattah, Nizar Saleh
    Dastiridou, Anna
    Sadda, SriniVas R.
    Lee, Olivia L.
    [J]. CORNEA, 2015, 34 : S48 - S52
  • [2] Clinical evaluation of the Oculus Keratograph
    Best, N.
    Drury, L.
    Wolffsohn, J. S.
    [J]. CONTACT LENS & ANTERIOR EYE, 2012, 35 (04) : 171 - 174
  • [3] RELIABILITY OF THE TEAR BREAK-UP TIME TECHNIQUE OF ASSESSING TEAR STABILITY AND THE LOCATIONS OF THE TEAR BREAK-UP IN HONG-KONG CHINESE
    CHO, P
    BROWN, B
    CHAN, I
    CONWAY, R
    YAP, M
    [J]. OPTOMETRY AND VISION SCIENCE, 1992, 69 (11) : 879 - 885
  • [4] Agreement between Automated and Traditional Measures of Tear Film Breakup
    Cox, Stephanie M.
    Nichols, Kelly K.
    Nichols, Jason J.
    [J]. OPTOMETRY AND VISION SCIENCE, 2015, 92 (09) : E257 - E263
  • [5] Foulks Gary N, 2003, Ocul Surf, V1, P20
  • [6] Noninvasive Assessment of Tear Stability with the Tear Stability Analysis System in Tear Dysfunction Patients
    Gumus, Koray
    Crockett, Charlene Hong
    Rao, Kavita
    Yeu, Elizabeth
    Weikert, Mitchell P.
    Shirayama, Mariko
    Hada, Shigeki
    Pflugfelder, Stephen C.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (01) : 456 - 461
  • [7] Izquierdo L, 2014, INVEST OPHTHALMOL VI, Vcitado 17 de agosto de 2016 55
  • [8] The effect of instilled fluorescein solution volume on the values and repeatability of TBUT measurements
    Johnson, ME
    Murphy, PJ
    [J]. CORNEA, 2005, 24 (07) : 811 - 817
  • [9] Comparison of fluorescein break-up time measurement reproducibility using standard fluorescein strips versus the Dry Eye Test (DET) method
    Korb, DR
    Greiner, JV
    Herman, J
    [J]. CORNEA, 2001, 20 (08) : 811 - 815
  • [10] KORB DR, 2002, TEAR FILM STRUCTURE, P9