A 256-element ultrasonic phased array system for the treatment of large volumes of deep seated tissue

被引:112
作者
Daum, DR [1 ]
Hynynen, K
机构
[1] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02215 USA
[2] Guidant Corp, St Paul, MN 55112 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/58.796130
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A 256-element phased array has been designed, constructed, and tested for ablative treatment of large focal volumes of deep seated tissue. The array was constructed fi om a 1.1-MHz, 1-3 composite piezoelectric spherical shell with a la-cm radius of curvature and a 12-cm diameter. The array was tested to determine its electroacoustic efficiency and inter-element coupling under high acoustic power conditions. A series of in vivo porcine experiments demonstrated the ability to produce deep seated tissue lesions in thigh muscle using the large scale phased array. The array was used to heat and coagulate tissue volumes > 5 cm(3) in a single ultrasound exposure using multiple foci and temporally scanned power deposition patterns. The spatial and temporal experimental results for large, heated focal volumes correlated very well with the simulated temperature response model for homogeneous tissue. A 25-cm(3) tissue volume was coagulated in a 90-min period using overlapping large ultrasound exposures.
引用
收藏
页码:1254 / 1268
页数:15
相关论文
共 46 条
[1]  
ANDERSON EM, 1995, ELECT TRANSMISSION L
[2]   HIGH-INTENSITY FOCUSED ULTRASOUND IN THE TREATMENT OF PROSTATIC TISSUE [J].
BIHRLE, R ;
FOSTER, RS ;
SANGHVI, NT ;
FRY, FJ ;
DONOHUE, JP .
UROLOGY, 1994, 43 (02) :21-26
[3]   CONCENTRIC-RING AND SECTOR-VORTEX PHASED-ARRAY APPLICATORS FOR ULTRASOUND HYPERTHERMIA [J].
CAIN, CA ;
UMEMURA, S .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :542-551
[4]  
Chung A, 1996, P SMR 4 M, P1751
[5]  
COLEMAN DJ, 1985, OPHTHALMOLOGY, V92, P339
[6]   FOCAL SPACING AND NEAR-FIELD HEATING DURING PULSED HIGH-TEMPERATURE ULTRASOUND THERAPY [J].
DAMIANOU, C ;
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :777-787
[7]   Design and evaluation of a feedback based phased array system for ultrasound surgery [J].
Daum, DR ;
Buchanan, MT ;
Fjield, T ;
Hynynen, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (02) :431-438
[8]  
DAUM DR, P 1997 IEEE ULTR S T
[9]  
DAUM DR, 1996 IEEE ULTR S P N, P1309
[10]   AMPLITUDE-DEPENDENT LOSSES IN ULTRASOUND EXPOSURE MEASUREMENT [J].
DUCK, FA ;
PERKINS, MA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :232-241