DENTAL AIR TURBINE HANDPIECE PERFORMANCE TESTING

被引:17
作者
DYSON, JE
DARVELL, BW
机构
[1] Department of Prosthetic Dentistry, University of Hong Kong
[2] Dental Materials Science Unit, University of Hong Kong
关键词
AIR TURBINE HANDPIECE; HIGH-SPEED EQUIPMENT; PERFORMANCE TESTING;
D O I
10.1111/j.1834-7819.1995.tb04823.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Air turbine handpieces are expected to continue to be widely used as the main means of carrying out dental cutting work and scope exists for further design improvements. An understanding of the theoretical principles governing the performance of these devices seems essential for the systematic development of better handpiece designs and methods of specification. Furthermore. for experimental work on cutting behaviour with air turbine equipment, this knowledge is required for appropriate characterization of the performance of the particular handpiece used with respect to actual rates of energy deposition. The literature relating to air turbine handpiece performance is critically reviewed to assess currently available methods of measuring important variables such as speed, torque, and power. In this, consideration is given to the current state of knowledge of the influence on these variables of air pressure, flow and turbine design features. It is apparent that, although various measurement methods have been described and data for individual handpieces published, no attempt has yet been made to explore the functional relationships that exist between the variables. It is concluded that there is a need to identify the factors influencing turbine performance, to develop measurement systems which would provide adequate accuracy and precision and then to investigate the functional relationships between these relevant variables.
引用
收藏
页码:330 / 338
页数:9
相关论文
共 53 条
[41]  
Miyari H, Nagai M, Fukuda H, Muramatsu A, Studies on the performance of the dental air turbine handpieces Part 3. Torque measurements of the air turbine handpieces, J Japan Soc Dent Appl Mater, 21, pp. 41-47, (1980)
[42]  
Fukuda H, Nagai M, Kobayashi M, Miyari H, Muramatsu A, Studies on the rotational performance of dental high‐speed handpieces, Reports Inst Med Dent Eng, 15, pp. 61-68, (1981)
[43]  
de Prony G, Note sur un moyen de mesurer l'effet dynamique des machines de rotation, Annal Chimie, 19, pp. 165-173, (1821)
[44]  
Jamieson A, A text‐book of heat and heat engines, (1919)
[45]  
Collett CV, Hope AD, Engineering measurements, (1983)
[46]  
Doebelin EO, Measurement systems—Application and design, (1983)
[47]  
Jamieson A
[48]  
Westland IAN, The energy requirement of the dental cutting process, J Oral Rehabil, 7, pp. 51-63, (1980)
[49]  
Bernier JL, Knapp MJ, Methods used in evaluation of high speed dental instruments and some results, Oral Surg Oral Med Oral Pathol, 12, pp. 234-240, (1959)
[50]  
Morrant GA, The use of air turbine handpieces in dentistry. I, Quintessence Int, 5, pp. 69-74, (1974)