Ten surgical patients were monitored with nasal and finger pulse oximetry (Nellcor N‐200) for five study periods with alternating mouth and nasal breathing and switching of cables and sensors. Nasal pulse oximetry was found to overestimate arterial oxygen saturation by 4.7 (SD 1.4%) (bias and precision). Copyright © 1990, Wiley Blackwell. All rights reserved