Comparative study between Air-Q and Intubating Laryngeal Mask Airway when used as conduit for fiber-optic

被引:17
作者
Abdel-Halim, Talaat M. [1 ]
El Enin, Mostafa A. Abo [1 ]
Elgoushi, Mohsen M. [1 ]
Afifi, Mohammed G. [1 ]
Atwa, Hesham S. [1 ]
机构
[1] Al Azhar Univ, Anaesthesia & Intens Care Dept, Cairo, Egypt
关键词
Air-Q; Fastrach LMA; FOB; Intubation; Conduit;
D O I
10.1016/j.egja.2013.10.004
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
This study compared Air-Q and Intubating LMA when used as a conduit for endotracheal intubation. Methods: One hundred patients scheduled for surgical operations under general anesthesia were randomly allocated into two equal groups (n=50). Group I: Air-Q and group II: Intubating Laryngeal Mask Airway (ILMA) in both groups intubation was done by Fiberoptic bronchoscope (FOB) through study device. After induction of anesthesia, patients were ventilated with Air-Q or ILMA. Then, endotracheal tube (ETT) was inserted through study device. Recorded measurements were as follows: number of attempts and duration of insertion of device, peak airway pressure and fiberoptic grading of laryngeal view. Also, we recorded number of attempts and duration of insertion of ETT and the incidence of blood stain on device and sore throat grading. Results: Duration of insertion of Air-Q was 13.300 +/- 3.471 s, whilst that of ILMA was 19.640 +/- 4.737 s (p < 0.001). In group I, peak airway pressure was 26.400 +/- 2.176 cmH(2)O, whilst, in group II, it was 25.260 +/- 1.468 cmH(2)O (p<0.01). Full view of vocal cords amounted to 78% and 26% of Groups I and II patients, respectively (p<0.001). Time of insertion of ETT was 33.5 +/- 6.795 s in group I, whilst in group II, it was 39.5 +/- 6.566 s (p<0.001). Blood stain was found on supraglottic device in 46% and 22% of cases in Groups I and II, respectively (p<0.01). Conclusion: Air-Q proved to be an excellent conduit for endotracheal intubation compared to the ILMA. (C) 2013 Production and hosting by Elsevier B.V. on behalf of Egyptian Society of Anesthesiologists. Open access under CC BY-NC-ND license.
引用
收藏
页码:107 / 113
页数:7
相关论文
共 18 条
[1]   Pilot study of the air-Q Intubating Laryngeal Airway in clinical use [J].
Bakker, E. J. ;
Valkenburg, M. ;
Galvin, E. M. .
ANAESTHESIA AND INTENSIVE CARE, 2010, 38 (02) :346-348
[2]   Air-Q the Intubating Laryngeal Airway: Comparative study of hemodynamic stress responses to tracheal intubation via Air-Q and direct laryngoscopy [J].
Bashandy, Ghada M. N. ;
Boules, Nermin S. .
EGYPTIAN JOURNAL OF ANAESTHESIA, 2012, 28 (02) :95-100
[3]   A proposed classification system for extraglottic airway devices [J].
Brimacombe, J .
ANESTHESIOLOGY, 2004, 101 (02) :559-559
[4]   The use of AIR-Q as conduit for fiberoptic endotracheal intubation in adult paralyzed patients [J].
Elganzouri, A. R. ;
Marzouk, Sahar ;
Youssef, Maha M. I. ;
Abdelalem, Nourhan .
EGYPTIAN JOURNAL OF ANAESTHESIA, 2012, 28 (04) :249-255
[5]   Preoperative airway assessment: Predictive value of a multivariate risk index [J].
ElGanzouri, AR ;
McCarthy, RJ ;
Tuman, KJ ;
Tanck, EN ;
Ivankovich, AD .
ANESTHESIA AND ANALGESIA, 1996, 82 (06) :1197-1204
[6]   CobraPLUS and Cookgas air-Q versus Fastrach for blind endotracheal intubation: a randomised controlled trial [J].
Erlacher, Wolfgang ;
Tiefenbrunner, Heide ;
Kaestenbauer, Thomas ;
Schwarz, Sylvia ;
Fitzgerald, Robert D. .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2011, 28 (03) :181-186
[7]   Use of the Intubating LMA-Fastrach™ in 254 Patients with Difficult-to-manage Airways [J].
Ferson, DZ ;
Rosenblatt, WH ;
Johansen, MJ ;
Osborn, I ;
Ovassapian, A .
ANESTHESIOLOGY, 2001, 95 (05) :1175-1181
[8]   The air-Q® intubating laryngeal airway vs the LMA-ProSeal™: a prospective, randomised trial of airway seal pressure [J].
Galgon, R. E. ;
Schroeder, K. M. ;
Han, S. ;
Andrei, A. ;
Joffe, A. M. .
ANAESTHESIA, 2011, 66 (12) :1093-1100
[9]   A Clinical Evaluation of the Intubating Laryngeal Airway as a Conduit for Tracheal Intubation in Children [J].
Jagannathan, Narasimhan ;
Kozlowski, Ryan J. ;
Sohn, Lisa E. ;
Langen, Kenneth E. ;
Roth, Andrew G. ;
Mukherji, Isabella I. ;
Kho, Melanie F. ;
Suresh, Santhanam .
ANESTHESIA AND ANALGESIA, 2011, 112 (01) :176-182
[10]   The second-generation air-Q® intubating laryngeal mask for airway maintenance during anaesthesia in adults: a report of the first 70 uses [J].
Joffe, A. M. ;
Liew, E. C. ;
Galgon, R. E. ;
Viernes, D. ;
Treggiari, M. M. .
ANAESTHESIA AND INTENSIVE CARE, 2011, 39 (01) :40-45