COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF A THYMIDYLATE SYNTHASE INHIBITOR, LY231514, AN INVESTIGATIONAL AGENT FOR THE TREATMENT OF SOLID TUMORS, IN HUMAN PLASMA

被引:13
作者
HAMILTON, CL [1 ]
KIRKWOOD, JA [1 ]
机构
[1] WISHARD MEM HOSP, LILLY LAB CLIN RES, INDIANAPOLIS, IN 46202 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 1994年 / 654卷 / 02期
关键词
D O I
10.1016/0378-4347(93)E0447-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A reversed-phase, column-switching high-performance liquid chromatographic (HPLC) method is described for the determination of a new thymidylate synthase inhibitor in human plasma. The compound and an internal standard are extracted from plasma using a Certify II solid-phase cartridge. Extracts are evaporated to dryness and the residue is reconstituted with mobile phase buffer. The analytes are separated from polar interferences and buffer salts originating from the elution step on a 4-mm YMC Basic pre-column. The fraction containing the analytes is further separated on a 25-cm YMC Basic column. The analytes are detected by their absorbance at 250 nm. The limit of quantitation is 10 ng/ml. The method is linear from 10 ng/ml to 80 mu g/ml using three standard curve ranges. Validation studies for all three ranges show the method to be reproducible. The method has been successfully used to support pharmacokinetic studies.
引用
收藏
页码:297 / 303
页数:7
相关论文
共 2 条
[1]  
HARRAP KR, 1989, ADV ENZYME REGUL, V29, P161, DOI 10.1016/0065-2571(89)90099-X
[2]  
JACKMAN AL, 1990, CANCER RES, V50, P5212