AN ABSOLUTE CRYOGENIC CALIBRATION SOURCE FOR A 2.5 GHZ RADIOMETER

被引:1
|
作者
GERVASI, M
BONELLI, G
SIRONI, G
CAVALIERE, F
PASSERINI, A
CASANI, S
机构
[1] CNR,IST IFCTR,I-20133 MILAN,ITALY
[2] UNIV MILAN,DIPARTIMENTO FIS,SEZIONE ASTROFIS,I-20133 MILAN,ITALY
[3] IST NAZL FIS NUCL,I-00185 ROME,ITALY
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1995年 / 66卷 / 10期
关键词
D O I
10.1063/1.1146155
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe a liquid helium-cooled absolute reference load for the calibration of a. 2.5 GHz radiometer. This calibration system provides an absolute reference near the liquid helium boiling point. The accuracy of the reference source is better than 100 mK. We use this cold load to perform absolute temperature measurements of the cosmic microwave background radiation. (C) 1995 American Institute of Physics.
引用
收藏
页码:4798 / 4805
页数:8
相关论文
共 50 条
  • [1] Design and calibration of electrical substitution radiometer traced to absolute cryogenic radiometer
    Zhang, Lei
    Xu, Wen-Bin
    Wang, Feng
    OPTICAL ENGINEERING, 2020, 59 (01)
  • [2] An absolute cryogenic radiometer for laser calibration and characterization of photodetectors
    Morozova, SP
    Konovodchenko, VA
    Sapritsky, VI
    Lisiansky, BE
    Morozov, PA
    Melenevsky, UA
    Petic, AG
    METROLOGIA, 1996, 32 (06) : 557 - 560
  • [3] Space Cryogenic Absolute Radiometer
    Wu D.
    Wang K.
    Ye X.
    Wang Y.-P.
    Fang W.
    Faguang Xuebao/Chinese Journal of Luminescence, 2019, 40 (08): : 1015 - 1021
  • [4] Prototype of Cryogenic Solar Absolute Radiometer and Transfer Radiometer for On-Board Calibration of Spectral Earth Imager
    Zajiczek, L.
    Winkler, R.
    Hobson, T.
    Green, P.
    Fox, N.
    13TH INTERNATIONAL CONFERENCE ON NEW DEVELOPMENTS AND APPLICATIONS IN OPTICAL RADIOMETRY, 2018, 972
  • [5] ABSOLUTE CALIBRATION OF AN ULTRAVIOLET SPECTROMETER USING A STABILIZED LASER AND A CRYOGENIC CAVITY RADIOMETER
    JAUNISKIS, L
    FOUKAL, P
    KOCHLING, H
    APPLIED OPTICS, 1992, 31 (27): : 5838 - 5843
  • [6] Experimental Study on Traceability of Absolute Radiometer to Cryogenic Radiometer
    Zhang Junchao
    Xu Nan
    Xu Ning
    Meng Haifeng
    Liu Wende
    Cai Chuan
    Zhang Bifeng
    Man Shuai
    Wang Meng
    Li Yinliang
    Xiong Limin
    ACTA OPTICA SINICA, 2022, 42 (05)
  • [7] A CRYOGENIC RADIOMETER FOR ABSOLUTE RADIOMETRIC MEASUREMENTS
    MARTIN, JE
    FOX, NP
    KEY, PJ
    METROLOGIA, 1985, 21 (03) : 147 - 155
  • [8] CRYOGENIC SOLAR ABSOLUTE RADIOMETER (CSAR)
    MARTIN, JE
    FOX, NP
    METROLOGIA, 1993, 30 (04) : 305 - 308
  • [9] Calibration of absolute spectral responsivity at 1 064 nm of transfer detector against cryogenic radiometer
    Pang, Weiwei
    Zheng, Xiaobing
    Li, Jianjun
    Shi, Xueshun
    Wu, Haoyu
    Xia, Maopeng
    Gao, Dongyang
    Shi, Jianmin
    Qi, Tao
    Kang, Qing
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2015, 44 (12): : 3812 - 3818
  • [10] A cryogenic radiometer for absolute neutron rate measurement
    Chowdhuri, Z
    Hansen, GL
    Jane, V
    Keith, CD
    Lozowski, WM
    Snow, WM
    Dewey, MS
    Gilliam, DM
    Greene, GL
    Nico, JS
    Thompson, AK
    Wietfeldtd, FE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (10): : 4280 - 4293