AN ACTIVE FEEDBACK-SYSTEM FOR ISOTONIC STUDIES OF SMOOTH-MUSCLE

被引:3
作者
SABOURIN, PJ
KINGMA, YJ
BOWES, KL
机构
[1] UNIV ALBERTA,DEPT ELECT ENGN,EDMONTON T6G 2N8,ALBERTA,CANADA
[2] UNIV ALBERTA,DEPT SURG,EDMONTON T6G 2N8,ALBERTA,CANADA
关键词
D O I
10.1109/10.81588
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A feedback system used to perform isotonic studies of smooth muscle is presented. This system is capable of applying a constant force to muscle samples regardless of their contractile activities. The force applied to the tissue is controlled using a proportional integral control system that drives a linear motor. The device is integrated into a sucrose gap tissue bath apparatus where measurements of displacement and electrical activity are also possible. The frequency of canine colonic smooth-muscle electrical oscillations is positively related to applied force.
引用
收藏
页码:614 / 616
页数:3
相关论文
共 8 条
[1]   MYOGENIC ELECTRICAL CONTROL ACTIVITY IN LONGITUDINAL MUSCLE OF HUMAN AND DOG COLON [J].
CHOW, E ;
HUIZINGA, JD .
JOURNAL OF PHYSIOLOGY-LONDON, 1987, 392 :21-34
[2]  
COBURN RF, METHOD PHARMACOL, V3, P231
[3]  
DEBSKI L, 1984, GASTROINTESTINAL MOT, P397
[4]   ELECTRICAL CORRELATE OF CIRCUMFERENTIAL CONTRACTIONS IN HUMAN COLONIC CIRCULAR MUSCLE [J].
HUIZINGA, JD ;
WATERFALL, WE .
GUT, 1988, 29 (01) :10-16
[5]  
OGATA K, 1970, MODERN CONTROL ENGIN, P216
[6]  
PEURA RA, 1978, MED INSTRUM, P49
[7]   SIMULTANEOUS MEASUREMENT OF ELECTRICAL-ACTIVITY FROM 2 COLONIC SMOOTH-MUSCLE LAYERS USING A DUAL SUCROSE GAP APPARATUS [J].
SABOURIN, PJ ;
KINGMA, YJ ;
BOWES, KL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (05) :509-514
[8]   ELECTRICAL AND MECHANICAL INTERACTIONS BETWEEN THE MUSCLE LAYERS OF CANINE PROXIMAL COLON [J].
SABOURIN, PJ ;
KINGMA, YJ ;
BOWES, KL .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 258 (03) :G484-G491