Clinical Validation of Computer-Assisted Navigation in Total Hip Arthroplasty

被引:5
作者
Beaumont, Eric [1 ]
Beaumont, Pierre [1 ]
Odermat, Daniel [2 ]
Fontaine, Isabelle [2 ]
Jansen, Herbert [2 ]
Prince, Francois [3 ]
机构
[1] Hop Sacre Coeur, Lab Rech Orthoped, 5400 Blvd Gouin Ouest,Locales K-3115, Montreal, PQ H4J 1C5, Canada
[2] ORTHO Soft Inc, Comp Assisted Surg, Montreal, PQ H3C 2N6, Canada
[3] Univ Montreal, Dept Kinesiol, Montreal, PQ H3T 1J4, Canada
关键词
D O I
10.4061/2011/171783
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A CT-based navigation system is helpful to evaluate the reamer shaft and the impactor position/ orientation during unilateral total hip arthroplasty (THA). The main objective of this study is to determine the accuracy of the Navitrack system by measuring the implant's true anteversion and inclination, based on pre-and postoperative CT scans (n = 9 patients). The secondary objective is to evaluate the clinical validity of measurements based on postop anteroposterior (AP) radiographs for determining the cup orientation. Postop CT-scan reconstructions and postop planar radiographs showed no significant differences in orientation compared to peroperative angles, suggesting a clinical validity of the system. Postoperative AP radiographs normally used in clinic are acceptable to determine the cup orientation, and small angular errors may originate from the patient position on the table.
引用
收藏
页数:6
相关论文
共 19 条
[1]   ANTEVERSION OF THE ACETABULAR CUP - MEASUREMENT OF ANGLE AFTER TOTAL HIP-REPLACEMENT [J].
ACKLAND, MK ;
BOURNE, WB ;
UHTHOFF, HK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :409-413
[2]   Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine [J].
Amiot, LP ;
Lang, K ;
Putzier, M ;
Zippel, H ;
Labelle, H .
SPINE, 2000, 25 (05) :606-614
[3]   The rationale for tilt-adjusted acetabular cup navigation [J].
Babisch, Juergen W. ;
Layher, Frank ;
Amiot, Louis-Philippe .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (02) :357-365
[4]  
Blendea Sorin, 2005, Comput Aided Surg, V10, P37, DOI 10.1080/10929080500178032
[5]  
BRIEN WW, 1993, ORTHOPEDICS, V16, P869
[6]  
CHANDLER DR, 1982, CLIN ORTHOP RELAT R, P284
[7]   The elevated-rim acetabular liner in total hip arthroplasty: Relationship to postoperative dislocation [J].
Cobb, TK ;
Morrey, BF ;
Ilstrup, DM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (01) :80-86
[8]   OPERATIVE CORRECTION OF AN UNSTABLE TOTAL HIP-ARTHROPLASTY [J].
DALY, PJ ;
MORREY, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (09) :1334-1343
[9]   Comparison of a mechanical acetabular alignment guide with computer placement of the socket [J].
DiGioia, AM ;
Jaramaz, B ;
Plakseychuk, AY ;
Moody, JE ;
Nikou, C ;
LaBarca, RS ;
Levison, TJ ;
Picard, F .
JOURNAL OF ARTHROPLASTY, 2002, 17 (03) :359-364
[10]   Orientation of the acetabular component - A comparison of five navigation systems with conventional surgical technique [J].
Honl, M. ;
Schwieger, K. ;
Salineros, M. ;
Jacobs, J. ;
Morlock, M. ;
Wimmer, M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (10) :1401-1405