Transfer of tactile input from an artificial hand to the forearm: experiments in amputees and able-bodied volunteers

被引:40
作者
Antfolk, Christian [1 ]
Cipriani, Christian [2 ]
Carrozza, Maria Chiara [2 ]
Balkenius, Christian [3 ]
Bjorkaman, Anders [4 ]
Lundborg, Goran [4 ]
Rosen, Birgitta [4 ]
Sebelius, Fredrik [1 ]
机构
[1] Lund Univ, Dept Measurement Technol & Ind Elect Engn, Lund, Sweden
[2] Scuola Super Sant Anna, BioRobot Inst, Pisa, Italy
[3] Univ Lund Cognit Sci, Lund, Sweden
[4] Skane Univ Hosp, Dept Hand Surg, Skane, Sweden
基金
瑞典研究理事会;
关键词
Hand prosthesis; haptic perception; sensory feedback;
D O I
10.3109/17483107.2012.713435
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Purpose: This study explores the possibilities of transferring peripheral tactile stimulations from an artificial hand to the forearm skin. Method: A tactile display applied to the forearm skin was used to transfer tactile input to the forearm from various locations on a hand displayed on a computer screen. Discernment of location, levels of pressure and a combination of the two in simulated functional grips was tested to quantify the participants' ability to accurately perceive the tactile stimulations presented. Ten participants (5 forearm amputees and 5 ablebodied volunteers) unfamiliar with the equipment participated in the three-stage experiments comprising a learning session with vision, a reinforced learning session without vision and a validation session without vision. Results: The location discernment accuracy was high in both groups (75.2% and 89.6% respectively). The capacity to differentiate between three different levels of pressure was also high (91.7% and 98.1% respectively in the two groups). Recognition of simulated grip was slightly more difficult with the groups scoring 58.7% and 68.0% respectively for accuracy in the validation session. Conclusions: This study demonstrates that it is possible, following a brief training period, to transfer tactile input from an artificial hand to the forearm skin. The level of accuracy was lower for the more complex task, simulated grip recognition, possibly because this represents a more complex task requiring higher order brain functions. These results could form the basis for developing sensory feedback in hand prostheses.
引用
收藏
页码:249 / 254
页数:6
相关论文
共 41 条
[1]   Design and technical construction of a tactile display for sensory feedback in a hand prosthesis system [J].
Antfolk, Christian ;
Balkenius, Christian ;
Lundborg, Goran ;
Rosen, Birgitta ;
Sebelius, Fredrik .
BIOMEDICAL ENGINEERING ONLINE, 2010, 9
[2]   SmartHand tactile display: A new concept for providing sensory feedback in hand prostheses [J].
Antfolk, Christian ;
Balkenius, Christian ;
Rosen, Birgitta ;
Lundborg, Goran ;
Sebelius, Fredrik .
JOURNAL OF PLASTIC SURGERY AND HAND SURGERY, 2010, 44 (01) :50-53
[3]  
Atkins DJ, 1996, J PROSTHET ORTHOT, V8, P2
[4]   Consumer design priorities for upper limb prosthetics [J].
Biddiss, Elaine ;
Beaton, Dorcas ;
Chau, Tom .
DISABILITY AND REHABILITATION-ASSISTIVE TECHNOLOGY, 2007, 2 (06) :346-357
[5]   Phantom digit somatotopy: a functional magnetic resonance imaging study in forearm amputees [J].
Bjorkman, Anders ;
Weibull, Andreas ;
Olsrud, Johan ;
Ehrsson, H. Henrik ;
Rosen, Birgitta ;
Bjorkman-Burtscher, Isabella M. .
EUROPEAN JOURNAL OF NEUROSCIENCE, 2012, 36 (01) :2098-2106
[6]  
Chappell P. H., 2011, Journal of Medical Engineering & Technology, V35, P1, DOI 10.3109/03091902.2010.526983
[7]  
Chatterjee A., 2008, JPO, V20, P27, DOI [10.1097/01.JPO.0000311041.61628.be, DOI 10.1097/01.JP0.0000311041.61628.BE]
[8]   On the shared control of an EMG-controlled prosthetic hand: Analysis of user-prosthesis interaction [J].
Cipriani, Christian ;
Zaccone, Franco ;
Micera, Silvestro ;
Carrozza, M. Chiara .
IEEE TRANSACTIONS ON ROBOTICS, 2008, 24 (01) :170-184
[9]   ON GRASP CHOICE, GRASP MODELS, AND THE DESIGN OF HANDS FOR MANUFACTURING TASKS [J].
CUTKOSKY, MR .
IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 1989, 5 (03) :269-279
[10]  
DARIO P, 1991, SENSOR ACTUAT A-PHYS, V26, P251, DOI 10.1016/0924-4247(91)87001-J