COMPETITIVE IMMUNOLOGICAL DETECTION OF PROGESTERONE BY MEANS OF THE ION-STEP INDUCED RESPONSE OF AN IMMUNOFET

被引:12
作者
SCHASFOORT, RBM [1 ]
KELDERMANS, CEJM [1 ]
KOOYMAN, RPH [1 ]
BERGVELD, P [1 ]
GREVE, J [1 ]
机构
[1] UNIV TWENTE,DEPT ELECT ENGN,7500 AE ENSCHEDE,NETHERLANDS
关键词
D O I
10.1016/0925-4005(90)80232-O
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The ion-step method was originally designed to measure immunoreactions via the change in charge density which occurs in an antibody loaded membrane, deposited on an ISFET, upon reaction with a charged antigen. In this paper we show the possibility of detecting an immunoreaction involving a non-charged antigen, progesterone, by using competitive binding of progesterone and a charged progesterone-lysozyme conjugate in the membrane. Both the immunoreaction and the non-specific binding could unequivocally be measured. The detection limit was approximately 10-8 M of progesterone in the sample solution. This proves that the ion-step method is suitable for detection of both charged and non-charged antigens. © 1990.