Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT

被引:0
作者
Wang, Hao [1 ]
Zhu, Li-Shuang [2 ]
Pang, Chen-Jiu [1 ]
Fan, Qi [1 ]
机构
[1] Henan Univ, Zhengzhou Univ, Henan Prov Peoples Hosp, Henan Eye Hosp,Peoples Hosp,Dept Refract Surg, 7 Weiwu Rd, Zhengzhou 450003, Peoples R China
[2] Zhengzhou Peoples Hosp, Dept Blood Transfus, 33 Huanghe Rd, Zhengzhou 450003, Peoples R China
关键词
OCT; Corneal thickness; Keratometry; Corneal volume; Repeatability; Agreement; OPTICAL COHERENCE TOMOGRAPHY; SCHEIMPFLUG CAMERA; PRECISION; BIOMETER;
D O I
10.1186/s12886-024-03448-z
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background The precision of anterior segment biometric measurements in eyes has become increasingly important in refractive surgery. The purpose of this study is to assess the repeatability of the automatic measurements provided by a new spectral-domain optical coherence tomograph (SD-OCT)/Placido topographer (MS-39, CSO) and its agreement with a swept-source OCT (SS-OCT) biometer (CASIA SS-1000, Tomey) in patients with myopia.Methods The right eye of 235 subjects was scanned 3 times with both devices. The evaluated parameters included central corneal radius of the steep meridian, central corneal radius of the flat meridian, mean central corneal radius, thinnest corneal thickness, central corneal thickness, anterior chamber depth, corneal volume and diameter. The intraobserver repeatability of the MS-39 measurements was calculated using intraclass correlation coefficient (ICC), within subject standard deviation, coefficient of repeatability, coefficient of variation and repeated-measures analysis of variance of the 3 repeated measurements. The agreement between the two devices was evaluated by 95% limits of agreement (LoA).Results The majority of the parameters acquired from MS-39 showed high repeatability. The repeatability of corneal diameter was slightly lower than the other measurements, although the ICC remained high. Agreement with the CASIA SS-1000 was good, indicated by the Bland-Altman plots with narrow 95% LoA values for all parameters assessed.Conclusions The high repeatability of automatic measurements by the new device supports its clinical application in eyes with myopia, and the good agreement between the two devices indicates they could be used interchangeably for the parameters evaluated.
引用
收藏
页数:9
相关论文
共 37 条
  • [1] Effects of diurnal, lighting, and angle-of-incidence variation on anterior segment optical coherence tomography (AS-OCT) angle metrics
    Akil, Handan
    Dastiridou, Anna
    Marion, Kenneth
    Francis, Brian A.
    Chopra, Vikas
    [J]. BMC OPHTHALMOLOGY, 2017, 17 : 31
  • [2] Dual versus single Scheimpflug camera for anterior segment analysis: Precision and agreement
    Aramberri, Jaime
    Araiz, Luis
    Garcia, Ane
    Illarramendi, Igor
    Olmos, Jaione
    Oyanarte, Izaskun
    Romay, Amaya
    Vigara, Itxaso
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (11) : 1934 - 1949
  • [3] Repeatability and agreement of swept-source optical coherence tomography and partial coherence interferometry biometers in myopes
    Ardoy, Elisa Garcia
    Mateos, Nuria
    Roda, Laura
    Sierra, Oscar Torrado
    Baptista, Antonio M.
    Serra, Pedro M.
    [J]. CLINICAL AND EXPERIMENTAL OPTOMETRY, 2023, 106 (07) : 783 - 792
  • [4] Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices
    Biswas, Sayantan
    Biswas, Partha
    [J]. OPTOMETRY AND VISION SCIENCE, 2021, 98 (10) : 1196 - 1202
  • [5] Measuring agreement in method comparison studies
    Bland, JM
    Altman, DG
    [J]. STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) : 135 - 160
  • [6] Improving precision for detecting change in the shape of the cornea in patients with keratoconus
    Brunner, Matthias
    Czanner, Gabriela
    Vinciguerra, Riccardo
    Romano, Vito
    Ahmad, Sajjad
    Batterbury, Mark
    Britten, Claire
    Willoughby, Colin E.
    Kaye, Stephen B.
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [7] Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik
    Chan, Tommy C. Y.
    Biswas, Sayantan
    Yu, Marco
    Jhanji, Vishal
    [J]. MEDICINE, 2015, 94 (30)
  • [8] Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group
    Chen, Min
    Zhang, Lina
    Xu, Jia
    Chen, Xinyi
    Gu, Yuxiang
    Ren, Yuping
    Wang, Kaijun
    [J]. BMC OPHTHALMOLOGY, 2019, 19 (01)
  • [9] Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
    Chen, Sisi
    Gao, Rongrong
    McAlinden, Colm
    Ye, Junming
    Wang, Yiran
    Chen, Min
    Huang, Jinhai
    Sun, Yong
    Yu, A-Yong
    [J]. JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [10] Swept Source Optical Coherence Tomography Analysis of the Selected Eye's Anterior Segment Parameters
    Dembski, Michal
    Nowinska, Anna
    Ulfik-Dembska, Klaudia
    Wylegala, Edward
    [J]. JOURNAL OF CLINICAL MEDICINE, 2021, 10 (05) : 1 - 13