Detector-based spectral CT with a novel dual-layer technology: principles and applications

被引:0
作者
Negin Rassouli
Maryam Etesami
Amar Dhanantwari
Prabhakar Rajiah
机构
[1] University Hospitals Cleveland Medical Center,Department of Radiology
[2] Yale University School of Medicine,Department of Radiology and Biomedical Imaging
[3] Philips Healthcare,Cardiothoracic Imaging, Department of Radiology
[4] UT Southwestern Medical Center,undefined
来源
Insights into Imaging | 2017年 / 8卷
关键词
CT; Dual-energy; Spectral; Contrast; Cardiac;
D O I
暂无
中图分类号
学科分类号
摘要
Detector-based spectral computed tomography is a novel dual-energy CT technology that employs two layers of detectors to simultaneously collect low- and high-energy data in all patients using standard CT protocols. In addition to the conventional polyenergetic images created for each patient, projection-space decomposition is used to generate spectral basis images (photoelectric and Compton scatter) for creating multiple spectral images, including material decomposition (iodine-only, virtual non-contrast, effective atomic number) and virtual monoenergetic images, on-demand according to clinical need. These images are useful in multiple clinical applications, including- improving vascular contrast, improving lesion conspicuity, decreasing artefacts, material characterisation and reducing radiation dose. In this article, we discuss the principles of this novel technology and also illustrate the common clinical applications.
引用
收藏
页码:589 / 598
页数:9
相关论文
共 64 条
[1]  
McCollough CH(2015)Dual and multienergy CT: principles, technical approaches, and clinical applications Radiology 276 637-653
[2]  
Leng S(2017)Spectral detector CT for cardiovascular applications Diagn Interv Radiol 23 187-193
[3]  
Yu L(1986)Evaluation of a prototype dual energy computed tomography apparatus. I. Phantom studies Med Phys 1986 13 334-339
[4]  
Fletcher JG(1976)Energy-selective reconstructions in X-ray computerized tomography Phys Med Biol 21 633-644
[5]  
Rajiah P(2012)Dual-energy CT: General Principles AJR Am J Roentgenol 199 S3-S8
[6]  
Abbara S(2011)Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner Med Phys 38 2222-2232
[7]  
Halliburton SS(2008)Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography Phys Med Biol 53 4031-4047
[8]  
Kalendar WA(2012)Dual-energy CT: vascular applications Am J Roengenol 199 587-597
[9]  
Perman WH(2014)Dual-energy CT of the abdomen Radiology 271 327-342
[10]  
Vetter JR(2014)Monoenergetic extrapolation of cardiac dual-energy CT for artifact reduction Acta Radiol 56 413-418