Dosimetric evaluation of a moving tumor target in intensity-modulated radiation therapy (IMRT) for lung cancer patients

被引:0
作者
Sung Kyu Kim
Min Kyu Kang
Ji Woon Yea
Se An Oh
机构
[1] Yeungnam University,Department of Radiation Oncology
[2] College of Medicine,Department of Physics
[3] Yeungnam University,Department of Radiation Oncology
[4] Yeungnam University Medical Center,undefined
来源
Journal of the Korean Physical Society | 2013年 / 63卷
关键词
Intensity-modulated radiation therapy; Movement of tumor position; Quality assurance; Dose verification;
D O I
暂无
中图分类号
学科分类号
摘要
Immobilization plays an important role in intensity-modulated radiation therapy (IMRT). The application of IMRT in lung cancer patients is very difficult due to the movement of the tumor target. Patient setup in radiation treatment demands high accuracy because IMRT employs a treatment size of a 1mm pixel unit. Hence, quality assurance of the dose delivered to patients must be at its highest. The radiation dose was evaluated for breathing rates of 9, 14, and 18 breaths per minute (bpm) for tumor targets moving up and down by 1.0 cm and 1.5 cm. The dose of the moving planned target volume (PTV) was measured by using a thermo-luminescent dosimeter (TLD) and Gafchromic™ EBT film. The measurement points were 1.0 cm away from the top, the bottom and the left and the right sides of the PTV center. The evaluated dose differences ranged from 94.2 to 103.8%, from 94.4 to 105.4%, and from 90.7 to 108.5% for 9, 14 and 18 bpm, respectively, for a tumor movement of 1.0 cm. The mean values of the doses were 101.4, 99.9, and 99.5% for 9, 14 and 18 bpm, respectively, for a tumor movement of 1.0 cm. Meanwhile, the evaluated dose differences ranged from 93.6 to 105.8%, from 95.9 to 111.5%, and from 96.2 to 111.7% for 9, 14 and 18 bpm, respectively, for a tumor movement of 1.5 cm. The mean values of the doses were 102.3, 103.4, and 103.1% for 9, 14 and 18 bpm, respectively, for a tumor movement of 1.5 cm. Therefore, we suggest that IMRT can be used in the treatment of lung cancer patients with vertical target movements within the range of 1.0 to 1.5 cm.
引用
收藏
页码:67 / 70
页数:3
相关论文
共 66 条
[1]  
Gierga D P(2004)undefined Int. J. Radiat. Oncol. Biol. Phys. 58 1584-undefined
[2]  
Chen G T Y(2003)undefined Int. J. Radiat. Oncol. Biol. Phys. 56 221-undefined
[3]  
Kung J H(2004)undefined Radiother. Oncol. 71 319-undefined
[4]  
Kitamura K(2003)undefined Int. J. Radiat. Oncol. Biol. Phys. 56 259-undefined
[5]  
Shirato H(2012)undefined J. Korean Phy. Soc. 60 1973-undefined
[6]  
Seppenwoolde Y(2003)undefined Radiot. & Oncol. 64 75-undefined
[7]  
Boehmer D(2001)undefined Cancer & Radiother. 5 472-undefined
[8]  
Bohsung J(2007)undefined Int. J. Radiat. Oncol. Biol. Phys. 68 1030-undefined
[9]  
Eichwurzel I(2007)undefined Int. J. Radiat. Oncol. Biol. Phys. 67 601-undefined
[10]  
Clippe S(2002)undefined J. Nucl. Med. 43 752-undefined