A probe and amplifier of a Q-meter for recording electron paramagnetic resonances at 300 MHz

被引:0
作者
A. E. Mefed
V. V. Demidov
机构
[1] Russian Academy of Sciences,Institute of Radio Engineering and Electronics, Fryazino Branch
[2] Russian Academy of Sciences,Institute of Radio Engineering and Electronics
来源
Instruments and Experimental Techniques | 2008年 / 51卷
关键词
07.57.Pt; 87.64.Hd;
D O I
暂无
中图分类号
学科分类号
摘要
A probe with a high-frequency amplifier of a continuous-wave Q-meter-type electron paramagnetic resonance (EPR) spectrometer operating at 300 MHz is described. The probe is inductively connected with both an exciting high-frequency generator and the amplifier. The amplifier is matched to the probe for obtaining a minimum noise figure of 2.9 dB. The single-scan EPR line with a 1.7-G width and 3 × 1018 resonant paramagnetic centers is recorded with a signal-to-noise ratio of about 260.