Cardiac index value measurement by invasive, semi-invasive and non invasive techniques: a prospective study in postoperative off pump coronary artery bypass surgery patients

被引:0
作者
Murali Chakravarthy
Subramanyam Rajeev
Vivek Jawali
机构
[1] Wockhardt Heart Institute,
来源
Journal of Clinical Monitoring and Computing | 2009年 / 23卷
关键词
cardiac index; arterial waveform; plethysmography; thermodilution; pulmonary artery; invasive; non invasive;
D O I
暂无
中图分类号
学科分类号
摘要
The authors prospectively compared three techniques of continuous cardiac index measurements. They were, invasive Continuous Cardiac Index using thermodilution flow-directed Swan-Ganz pulmonary artery catheters, semi invasive Flotrac™—arterial pressure derived cardiac index and the non invasive cardiac index measurement—body impedance plethsmography. The cardiac index measurements were made simultaneously in the postoperative period in 20 patients who underwent elective uncomplicated off pump coronary artery bypass graft. The values were collected once in 5 min over a period of 30–40 min. A set of 140 values were obtained from the cohorts in our study. Inter-changeability of the values of cardiac index was analysed using Bland-Altman and mountain plots. The cardiac index values ranged from 1.6 to 3.6 l/min/m2. The values obtained were interchangeable. The bias and precision respectively were 0.02 and ±0.06 for continuous cardiac index and Flotrac™, 0.18 and ±0.08 for Flotrac™ and body impedance plethysmography and 0.16 and ±0.08 for continuous cardiac index and body impedance plethysmography. Flotrac™ appears to be more useful during off pump coronary artery bypass surgery.
引用
收藏
页码:175 / 180
页数:5
相关论文
共 66 条
[1]  
Chakravarthy M(2008)Cardiac output–have we found the ‹gold standard’? Ann Card Anaesth 11 1-2
[2]  
Mayer J(2007)Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery Br J Anaesth 98 176-182
[3]  
Boldt J(2001)Pulmonary artery catheter monitoring does not reduce morbidity and or mortality in major surgery Ugeskr Laeger 163 5212-295
[4]  
Schöllhorn T(2006)Swan-Ganz catheter-induced severe complications in cardiac surgery: right ventricular perforation, knotting, and rupture of a pulmonary artery J Card Surg 21 292-146
[5]  
Röhm KD(2007)Pro: Is the pulmonary artery catheter dead? J Cardiothorac Vasc Anesth 21 144-80
[6]  
Mengistu AM(2006)Noninvasive, automated and continuous cardiac output monitoring by pulmonary capnodynamics: breath-by-breath comparison with ultrasonic flow probe Anesthesiology 105 72-620
[7]  
Suttner S(1985)The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model J Biomech 18 613-310
[8]  
Larsson A(1986)Statistical methods for assessing agreement between two methods of clinical measurement Lancet 1 307-527
[9]  
Bossert T(1995)A simple, graphical method to evaluate laboratory assays Eur J Clin Chem Clin Biochem 33 525-7
[10]  
Gummert JF(2007)Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration J Cardiothorac Vasc Anesth 21 3-126