Accuracy validation of incident photon fluence on DQE for various measurement conditions and X-ray units

被引:4
作者
Haba T. [1 ,2 ]
Kondo S. [1 ]
Hayashi D. [1 ]
Koyama S. [1 ]
机构
[1] Department of Radiological Technology, Graduate School of Medicine, Nagoya University, Higashi-ku, Nagoya
[2] Department of Radiology, Fujita Health University Hospital, Toyoake, Aichi, 1-98 Dengakugakubo, Kutsukake-cho
关键词
Detective quantum efficiency; Incident photon fluence; X-ray spectrum;
D O I
10.1007/s12194-013-0215-4
中图分类号
学科分类号
摘要
Detective quantum efficiency (DQE) is widely used as a comprehensive metric for X-ray image evaluation in digital X-ray units. The incident photon fluence per air kerma (SNRin2) is necessary for calculating the DQE. The International Electrotechnical Commission (IEC) reports the SNR in2 under conditions of standard radiation quality, but this SNRin2 might not be accurate as calculated from the X-ray spectra emitted by an actual X-ray tube. In this study, we evaluated the error range of the SNRin2 presented by the IEC62220-1 report. We measured the X-ray spectra emitted by an X-ray tube under conditions of standard radiation quality of RQA5. The spectral photon fluence at each energy bin was multiplied by the photon energy and the mass energy absorption coefficient of air; then the air kerma spectrum was derived. The air kerma spectrum was integrated over the whole photon energy range to yield the total air kerma. The total photon number was then divided by the total air kerma. This value is the SNRin2. These calculations were performed for various measurement parameters and X-ray units. The percent difference between the calculated value and the standard value of RQA5 was up to 2.9 %. The error range was not negligibly small. Therefore, it is better to use the new SNRin2 of 30694 (1/(mm2 μGy)) than the current SNRin2 of 30174 (1/(mm2 μGy)). © 2013 Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:423 / 430
页数:7
相关论文
共 10 条
[1]  
Dainty J.C., Shaw R., Image Science, (1974)
[2]  
Medical Electrical Equipment - Characteristics of Digital X-ray Imaging Devices, (2008)
[3]  
Kramer H., SPEVAL Software Package Version, (1995)
[4]  
Di Castro E., Pani R., Pellegrini R., Bacci C., The use of cadmium telluride detectors for the qualitative analysis of diagnostic X-ray spectra, Physics in Medicine and Biology, 29, 9, pp. 1117-1131, (1984)
[5]  
Matsumoto C., Performance of a new schottky CdTe detector for hard x-ray spectroscopy, IEEE Transactions on Nuclear Science, 45, 3 PART 1, pp. 428-432, (1998)
[6]  
Takahashi T., Mitani T., Kobayashi Y., Kouda M., Sato G., Watanabe S., Nakazawa K., Okada Y., Funaki M., Ohno R., Mori K., High-resolution Schottky CdTe diode detector, IEEE Transactions on Nuclear Science, 49, 3, pp. 1297-1303, (2002)
[7]  
Miyajima S., Thin CdTe detector in diagnostic x-ray spectroscopy, Medical Physics, 30, 5, pp. 771-777, (2003)
[8]  
Hubbell J.H., Seltzer S.M., Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients from 1 KeV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest, (1996)
[9]  
Tucker D.M., Semiempirical model for generating tungsten target x-ray spectra, Med Phys, 18, pp. 211-218, (1991)
[10]  
Medical Diagnostic X-ray Equipment - Radiation Conditions for Use in the Determination of Characteristics, (2005)