Equivalent Size and Acoustic Center of Measuring Hydrophone

被引:0
作者
A. E. Isaev
B. I. Khatamtaev
机构
[1] Russian Metrological Institute of Technical Physics and Radio Engineering,
来源
Measurement Techniques | 2023年 / 65卷
关键词
hydrophone; amplitude-phase hydrophone calibration; reciprocity method; hydrophone acoustic center; active element equivalent radius;
D O I
暂无
中图分类号
学科分类号
摘要
The article considers the problem that arises during the amplitude-phase calibration of hydrophones by the reciprocity method and consists in the violation of the consistency of the hydrophone characteristics specified by the manufacturer. Violation of the indicated consistency of characteristics is due to the need to determine the phase-frequency characteristic when calibrating the hydrophone. The manufacturer sets the dimensions of the hydrophone receiving part and the position of the acoustic center according to the drawing of the receiving part and thereby replaces the acoustic parameters with geometric ones. To restore the consistency of the characteristics, it is proposed to determine the position of the acoustic center and the equivalent size of the hydrophone receiving part from the results of acoustic measurements. It is shown that the displacement of the hydrophone acoustic center relative to its geometric center is a typical phenomenon, and the geometric and acoustically measured dimensions of the active element can differ. To ensure the unity of acoustic measurements, it is proposed to introduce the concept of an equivalent radius and to clarify the concept of the hydrophone acoustic center, as well as to standardize the methods of their experimental determining. The results obtained will be useful in monitoring the phase responses of elements of hydroacoustic antennas for various purposes, ensuring the identity of the phase-frequency characteristics of the vector receivers measuring channels, in solving problems related to restoring the acoustic signal shape by the recorded voltage at the receiver output and other various applications, including marine ecology, medical ultrasound studies, etc.
引用
收藏
页码:929 / 935
页数:6
相关论文
共 31 条
  • [1] Bazulin EG(2019)Reconstruction of Ultrasound Reflector Images from Incomplete Data using the Compressive Sensing Method Acoust. Phys. 65 432-443
  • [2] Sokolov DM(2020)Designing fully Populated Phased Arrays for Noninvasive Ultrasound Surgery with Controlled Degree of Irregularity in the Arrangement of Elements Acoust. Phys. 66 352-361
  • [3] Rosnitskiy PB(2021)Equations for Calculating the Amplitude-Frequency and Phase-Frequency Responses of a Tripole-Type Vector-Scalar Receiver with a Time Delay of a Monopole Signal Acoust. Phys. 67 421-429
  • [4] Sapozhnikov OA(2008)A Nonlinear Propagation Model-Based Phase Calibration Technique for Membrane Hydrophones IEEE T. Ultrason. Ferr. 55 84-93
  • [5] Gavrilov LR(2011)Time-Delay Spectrometry Measurement of Magnitude and Phase of Hydrophone Response IEEE T. Ultrason. Ferr. 58 2325-2333
  • [6] Khokhlova VA(1981)Phase Calibration of Hydrophones J. Acoust. Soc. Am. 70 516-519
  • [7] Kuznetsov GN(2013)Measurement of the Sensitivity Phase-Frequency Characteristics of Hydrophones by the Reciprocity Method Meas. Tech. 56 706-711
  • [8] Pudovkin AA(2021)Determination of the Hydrophone Phase-Frequency Response by Its Amplitude-Frequency Response Meas. Tech. 64 580-585
  • [9] Subbotkin AO(2022)Experimental Determination of the Acoustic Center of a Measuring Hydrophone Al'manakh Sovr. Metrol. 30 56-71
  • [10] Cooling MP(2008)Difference between Nominal and Measured Active Element Sizes of Hydrophones Jpn. J. Appl. Phys. 47 3926-3928