Optimisation of multislice computed tomography protocols in angio-CT examinations

被引:3
作者
Strocchi S. [1 ]
Vite C. [1 ]
Callegari L. [2 ]
Conte L. [3 ]
机构
[1] Fisica Sanitaria, Ospedale di Circolo e Fondazione Macchi, I-21100 Varese
[2] Radiologia, Ospedale di Circolo e Fondazione Macchi, I-21100 Varese
[3] Dipartimento di Scienze Cliniche e Biologiche, Università dell'Insubria
关键词
Effective dose; Multislice computed tomography;
D O I
10.1007/s11547-006-0024-5
中图分类号
学科分类号
摘要
Purpose. The aim of this paper is to explain a general procedure for the optimisation of multislice computed tomography (MSCT) protocols. Materials and methods. Four angio-CT protocols with a GE LightSpeed Plus 4-slice CT scanner were considered. Effective doses were computed for a sample of patients. First the dose was optimised for arterial-phase scans on a standard patient and adapted to the weight of individual patients with a scaling factor. Results. The mean effective dose for an angio-CT examination ranged from 18.8 mSv to 28.8 mSv, depending on the protocol adopted. Following the optimisation procedure, we drew up a table indicating tube current values for each patient weight. Calculation of the effective dose before and after the optimisation procedure revealed a dose reduction of about 40%. Conclusions. Angio-CT examinations deliver high doses, but these doses can be reduced without affecting image quality.
引用
收藏
页码:238 / 244
页数:6
相关论文
共 8 条
  • [1] Jones D.G., Shrimpton P.C., Survey of CT Practice in the UK. Part 2: Dosimetric Aspects, (1991)
  • [2] Jones D.G., Shrimpton P.C., Survey of CT Practice in the UK Part 3: Normalized Organ Doses for X-ray Computed Tomography Calculated Using Monte Carlo Tecniques, (1991)
  • [3] Radiation Dosimetry: X-rays Generated at Potentials of 5 to 150 KV, (1970)
  • [4] Reccomendation of the International Commision on Radiological Protection, (1990)
  • [5] Wilting J.E., Zwartkruis A., Vanleeuwen M.S., Et al., A rational approach to dose reduction in CT: Individualized scan protocols, Eur Radiol, 11, pp. 2627-2632, (2001)
  • [6] Boone J.M., Dose reduction in pediatric CT: A rational approach, Radiology, 228, pp. 352-360, (2003)
  • [7] Brooks R.A., Di Chiro G., Statistical limitation in X-ray recostructive tomography, Med Phys, 3, pp. 273-1240, (1976)
  • [8] Nickoloff E.L., Dutta A.K., Influence of phantom diameter, kVp and scan mode upon computed tomography index, Med Phys, 30, pp. 395-402, (2003)