Attenuation correction of myocardial SPECT images with X-ray CT: Effects of registration errors between X-ray CT and SPECT

被引:0
作者
Yasuyuki Takahashi
Kenya Murase
Hiroshi Higashino
Teruhito Mochizuki
Nobutoku Motomura
机构
[1] Osaka Graduate School of Medicine,Department of Medical Engineering, Division of Allied Health Sciences
[2] Ehime University School of Medicine,Department of Radiology
[3] Toshiba Medical Engineering Laboratory,Department of Radiology
[4] Ehime Prefectural Imabari Hospital,undefined
来源
Annals of Nuclear Medicine | 2002年 / 16卷
关键词
SPECT; X-ray CT; fusion image; misaligned images;
D O I
暂无
中图分类号
学科分类号
摘要
Purpose: Attenuation correction with an X-ray CT image is a new method to correct attenuation on SPECT imaging, but the effect of the registration errors between CT and SPECT images is unclear. In this study, we investigated the effects of the registration errors on myocardial SPECT, analyzing data from a phantom and a human volunteer.Methods: Registerion (fusion) of the X-ray CT and SPECT images was done with standard packaged software in three dimensional fashion, by using linked transaxial, coronal and sagittal images. In the phantom study, an X-ray CT image was shifted 1 to 3 pixels on thex, y andz axes, and rotated 6 degrees clockwise. Attenuation correction maps generated from each misaligned X-ray CT image were used to reconstruct misaligned SPECT images of the phantom filled with201Tl. In a human volunteer, X-ray CT was acquired in different conditions (during inspiration vs. expiration). CT values were transferred to an attenuation constant by using straight lines; an attenuation constant of 0/cm in the air (CT value=−1,000 HU) and that of 0.150/cm in water (CT value=0 HU). For comparison, attenuation correction with transmission CT (TCT) data and an external γ-ray source (99mTc) was also applied to reconstruct SPECT images.Results: Simulated breast attenuation with a breast attachment, and inferior wall attenuation were properly corrected by means of the attenuation correction map generated from X-ray CT. As pixel shift increased, deviation of the SPECT images increased in misaligned images in the phantom study. In the human study, SPECT images were affected by the scan conditions of the X-ray CT.Conclusion: Attenuation correction of myocardial SPECT with an X-ray CT image is a simple and potentially beneficial method for clinical use, but accurate registration of the X-ray CT to SPECT image is essential for satisfactory attenuation correction.
引用
收藏
页码:431 / 435
页数:4
相关论文
共 70 条
[1]  
Murase K(1993)Improvement of brain single photon emission tomography (SPET) using transmission data acquisition in a four-head SPET scanner Eur J Nucl Med 20 32-38
[2]  
Tanada S(1998)Dosimetry of transmission measurements in nuclear medicine: a study using anthromorphic phantoms and thermoluminescent dosimeters Eur J Nucl Med 25 1435-1441
[3]  
Inoue T(2000)Gamma camera-mounted anatomical X-ray tomography: technology, system characteristics and first images Eur J Nucl Med 27 619-627
[4]  
Sugawara Y(2000)Image fusion using an integrated, dual-head coincidence camera with X-ray tube-based attenuation maps J Nucl Med 41 1364-1368
[5]  
Hamamoto K(1997)Myocardial perfusion imaging with a combined x-ray CT and SPECT system J Nucl Med 38 1535-1540
[6]  
Almeida P(2001)Receiver operating characteristic (ROC) analysis of image reconstructed with iterative expectation maximization algorithms Ann Nucl Med 15 521-525
[7]  
Bendriem B(1999)Examination of the SPECT attenuation correction using the X-ray CT image KAKU IGAKU (Jpn J Nucl Med) 36 617-617
[8]  
Dreuille O(1995)A Fully Automatic Multimodality Image Registration Algorithm J Comput Assist Tomogr 19 615-623
[9]  
Peltier A(1989)A technique for using CT images in attenuation correction and quantification in SPECT Nucl Med Commun 10 83-97
[10]  
Perrot C(2002)CT vs Eur J Nucl Med 29 346-350