Intraocular pressure control after the implantation of a second Ahmed glaucoma valve

被引:0
|
作者
Jesús Jiménez-Román
Félix Gil-Carrasco
Vital Paulino Costa
Rui Barroso Schimiti
Fabián Lerner
Priscila Rezende Santana
Jose Paulo Cabral Vascocellos
Armando Castillejos-Chévez
Mauricio Turati
Karina Fabre-Miranda
机构
[1] Asociación Para Evitar la Ceguera en México,Glaucoma Department
[2] UNICAMP,Glaucoma Department
[3] Fundación para el estudio del Glaucoma,Glaucoma Department
来源
International Ophthalmology | 2016年 / 36卷
关键词
Glaucoma drainage device; Ahmed drainage device; Glaucoma; Surgery;
D O I
暂无
中图分类号
学科分类号
摘要
The objective of this study is to evaluate the efficacy and safety of a second Ahmed glaucoma valve (AGV) in eyes with refractory glaucoma that had undergone prior Ahmed device implantation. This multicenter, retrospective study evaluated 58 eyes (58 patients) that underwent a second AGV (model S2-n = 50, model FP7-n = 8) due to uncontrolled IOP under maximal medical therapy. Outcome measures included IOP, visual acuity, number of glaucoma medications, and postoperative complications. Success was defined as IOP <21 mmHg (criterion 1) or 30 % reduction of IOP (criterion 2) with or without hypotensive medications. Persistent hypotony (IOP <5 mmHg after 3 months of follow-up), loss of light perception, and reintervention for IOP control were defined as failure. Mean preoperative IOP and mean IOPs at 12 and 30 months were 27.55 ± 1.16 mmHg (n = 58), 14.45 ± 0.83 mmHg (n = 42), and 14.81 ± 0.87 mmHg (n = 16), respectively. The mean numbers of glaucoma medications preoperatively at 12 and 30 months were 3.17 ± 0.16 (n = 58), 1.81 ± 0.2 (n = 42), and 1.83 ± 0.35 (n = 18), respectively. The reductions in mean IOP and number of medications were statistically significant at all time intervals (P < 0.001). According to criterion 1, Kaplan–Meier survival curves disclosed success rates of 62.9 % at 12 months and 56.6 % at 30 months. According to criterion 2, Kaplan–Meier survival curves disclosed success rates of 43.9 % at 12 months and 32.9 % at 30 months. The most frequent early complication was hypertensive phase (10.3 %) and the most frequent late complication was corneal edema (17.2 %). Second AGV implantation may effectively reduce IOP in eyes with uncontrolled glaucoma, and is associated with relatively few complications.
引用
收藏
页码:347 / 353
页数:6
相关论文
共 50 条
  • [1] Intraocular pressure control after the implantation of a second Ahmed glaucoma valve
    Jimenez-Roman, Jesus
    Gil-Carrasco, Felix
    Costa, Vital Paulino
    Schimiti, Rui Barroso
    Lerner, Fabian
    Santana, Priscila Rezende
    Cabral Vascocellos, Jose Paulo
    Castillejos-Chevez, Armando
    Turati, Mauricio
    Fabre-Miranda, Karina
    INTERNATIONAL OPHTHALMOLOGY, 2016, 36 (03) : 347 - 353
  • [2] Ahmed Glaucoma Valve Implantation to Reduce Intraocular Pressure: Updated Perspectives
    Arikan, Gul
    Dokuz, Uzeyir Gunenc
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 1833 - 1845
  • [3] Evaluation of success after second Ahmed glaucoma valve implantation
    Nilforushan, Naveed
    Yadgari, Maryam
    Jazayeri, Anis Alsadat
    Karimi, Nasser
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2016, 64 (03) : 206 - 210
  • [4] Ahmed valve implantation to control intractable glaucoma after phacoemulsification and intraocular lens implantation in a dog
    García-Sánchez, GA
    Whitley, RD
    Brooks, DE
    Trigo, F
    Piñón, A
    VETERINARY OPHTHALMOLOGY, 2005, 8 (02) : 139 - 144
  • [5] Effect of immediate shallow anterior chamber after Ahmed glaucoma valve implantation on intermediate-term intraocular pressure control
    Patil, Trupti Sudhir
    Mani, Baskaran
    Balekudaru, Shantha
    George, Ronnie J.
    Lingam, Vijaya
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2022, 70 (08) : 2915 - 2921
  • [6] Non-intraocular pressure-related revision surgeries after Ahmed glaucoma valve (AGV) implantation in refractory glaucoma
    Asli Kirmaci Kabakci
    Banu Solmaz
    Berna Basarir
    Burcu Kemer Atik
    Tekin Yasar
    International Ophthalmology, 2021, 41 : 3533 - 3538
  • [7] Non-intraocular pressure-related revision surgeries after Ahmed glaucoma valve (AGV) implantation in refractory glaucoma
    Kabakci, Asli Kirmaci
    Solmaz, Banu
    Basarir, Berna
    Atik, Burcu Kemer
    Yasar, Tekin
    INTERNATIONAL OPHTHALMOLOGY, 2021, 41 (10) : 3533 - 3538
  • [8] The Ahmed Glaucoma Valve - a new implant for high intraocular pressure
    Hille, K
    Ruprecht, KW
    OPHTHALMOLOGE, 1999, 96 (10): : 653 - 657
  • [9] Surgical Outcomes of Additional Ahmed Glaucoma Valve Implantation in Refractory Glaucoma
    Ko, Sung Ju
    Hwang, Young Hoon
    Ahn, Sang Il
    Kim, Hwang Ki
    JOURNAL OF GLAUCOMA, 2016, 25 (06) : E620 - E624
  • [10] Ocular massage for intraocular pressure control after Ahmed valve insertion
    McIlraith, Ian
    Buys, Yvonne
    Campbell, Robert J.
    Trope, Graham E.
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2008, 43 (01): : 48 - 52