Development of a novel shrouded impeller pediatric blood pump

被引:9
作者
Khan, Talha Irfan [1 ]
Zad, Haris Sheh [1 ]
Lazoglu, Ismail [1 ]
Yalcin, Ozlem [2 ]
机构
[1] Koc Univ, Mfg & Automat Res Ctr, TR-34450 Istanbul, Turkey
[2] Koc Univ, Sch Med, Istanbul, Turkey
关键词
Pediatric ventricular assist device; Computational fluid dynamics; Shrouded impeller; Hydraulic performance; VENTRICULAR ASSIST DEVICE; CHILDREN; SUPPORT; DAMAGE;
D O I
10.1007/s10047-018-1028-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this work was to analyze a shrouded impeller pediatric ventricular assist device (SIP-VAD). This device has distinctive design characteristics and parameter optimizations for minimization of recirculation flow and reduction in high-stress regions that cause blood damage. Computational Fluid Dynamics (CFD) simulations were performed to analyze the optimized design. The bench-top prototype of SIP-VAD was manufactured with biocompatible stainless steel. A study on the hydrodynamic and hemodynamic performance of the SIP-VAD was conducted with predictions from CFD and actual experimentation values, and these results were compared. The CFD analysis yielded a pressure range of 29-90 mmHg corresponding to flow rates of 0.5-3 L/min over 9000-11000 rpm. The predicted value of the normalized index of hemolysis (NIH) was 0.0048 g/100 L. The experimental results with the bench-top prototype showed a pressure rise of 30-105 mmHg for the flow speed of 8000-12000 rpm and flow rate of 0.5-3.5 L/min. The maximum difference between CFD and experimental results was 4 mmHg pressure. In addition, the blood test showed the average NIH level of 0.00674 g/100 L. The results show the feasibility of shrouded impeller design of axial-flow pump for manufacturing the prototype for further animal trials.
引用
收藏
页码:142 / 149
页数:8
相关论文
共 17 条
[1]   Waiting List Mortality Among Children Listed for Heart Transplantation in the United States [J].
Almond, Christopher S. D. ;
Thiagarajan, Ravi R. ;
Piercey, Gary E. ;
Gauvreau, Kimberlee ;
Blume, Elizabeth D. ;
Bastardi, Heather J. ;
Fynn-Thompson, Francis ;
Singh, T. P. .
CIRCULATION, 2009, 119 (05) :717-727
[2]  
Baldwin JT, 2017, SEMINARS THORACIC CA, P9
[3]   A review of computational fluid dynamics analysis of blood pumps [J].
Behbahani, M. ;
Behr, M. ;
Hormes, M. ;
Steinseifer, U. ;
Arora, D. ;
Coronado, O. ;
Pasquali, M. .
EUROPEAN JOURNAL OF APPLIED MATHEMATICS, 2009, 20 :363-397
[4]  
Benjamin EJ, 2018, CIRCULATION, V137, pE67, DOI [10.1161/CIR.0000000000000558, 10.1161/CIR.0000000000000485, 10.1161/CIR.0000000000000530]
[5]   The PediPump: A new ventricular assist device for children [J].
Duncan, BW ;
Lorenz, M ;
Kopcak, MW ;
Fukamachi, K ;
Ootaki, Y ;
Chen, HM ;
Chapman, PA ;
Davis, SJ ;
Smith, WA .
ARTIFICIAL ORGANS, 2005, 29 (07) :527-530
[6]   Mechanical circulatory support for infants and children with cardiac disease [J].
Duncan, BW .
ANNALS OF THORACIC SURGERY, 2002, 73 (05) :1670-1677
[7]   Fast three-dimensional numerical hemolysis approximation [J].
Garon, A ;
Farinas, MI .
ARTIFICIAL ORGANS, 2004, 28 (11) :1016-1025
[8]   Strain-based blood damage estimation for computational design of ventricular assist devices [J].
Gesenhues, Linda ;
Pauli, Lutz ;
Behr, Marek .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2016, 39 (04) :166-170
[9]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306
[10]   Principle and basic property of the sequential flow pump [J].
Hara, Shintaro ;
Maeno, Erina ;
Li, Xinyang ;
Yurimoto, Terumi ;
Isoyama, Takashi ;
Saito, Itsuro ;
Ono, Toshiya ;
Abe, Yusuke .
JOURNAL OF ARTIFICIAL ORGANS, 2017, 20 (03) :215-220