Use of through-and-through guidewire for delivering large stent-grafts into the distal aortic arch
被引:0
|作者:
Muhammad Al Shammari
Peter Taylor
John F. Reidy
机构:
[1] Guy’s Hospital,Department of Radiology, 2nd Floor, Guy’s Tower
[2] Guy’s Hospital,Department of Surgery, 2nd Floor New Guy’s House
来源:
CardioVascular and Interventional Radiology
|
2000年
/
23卷
关键词:
Aorta, aneurysm;
Stents and prostheses;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
The availability of large diameter stent-grafts is now allowing the endovascular treatment of thoracic aortic aneurysms. Most aneurysms are closely related to the distal arch and it is thus necessary to pass the delivery systems into the arch to effectively cover the proximal neck. Even with extra-stiff guidewires in position, it may still be difficult to achieve this, as a result of tortuosity at the iliac arteries and the aorta. We detail a technique where a stiff guidewire is passed from a brachial entry point through the aorta and out at the femoral arteriotomy site. This allows extra-support and may enable the delivery system to be passed further into the aortic arch than it could with just the regular guidewire position.