Estimation of Reference Gamma Radiation Field for Calibration of Radiation Monitors

被引:0
|
作者
Sunil K. Singh
Liji Shaiju
Aashna Gupta
S. M. Tripathi
V. Sathian
Probal Chaudhury
机构
[1] Bhabha Atomic Research Centre,Radiation Safety Systems Division
[2] Department of Atomic Energy,undefined
来源
MAPAN | 2024年 / 39卷
关键词
Air kerma; Air kerma strength; Calibration; Radiation monitor; Inverse square law;
D O I
暂无
中图分类号
学科分类号
摘要
Laboratories/facilities dealing with ionizing radiation encounter radiation field which are measured using suitable gamma radiation monitors, for area and personal monitoring. These radiation monitors are subjected to periodic calibrations to confirm that they are still fit for the purpose. In India, periodic calibration of these radiation monitors is carried out by NABL accredited calibration laboratories using source exposure devices containing radionuclide sources of 137Cs and/or 60Co. Calibration laboratories get the output from these source exposure devices standardized by RSSD, BARC, Mumbai. RSSD, BARC, provides a report on the measured output, in the form of air kerma rates at different distances, which is traceable to relevant national standards. The laboratory then uses the standardized radiation field in day-to-day calibration, by taking air kerma rate mostly at distance of 100 cm (fixed distance) as reference for estimating radiation fields at other distances by using inverse square law. The air kerma rate values obtained by the using inverse square law shows deviation from the measured values with an allowed deviation up to 5%. The impact of this deviation cannot be corrected thus is accounted in the form of uncertainty. This paper presents an average air kerma strength-based methodology which provides a simple and effective mechanism for estimating the conventional true value of air kerma rate at various distance which adds lower uncertainty to overall uncertainty, as compared to the method of using a fixed or reference distance (say 100 cm) based air kerma rate.
引用
收藏
页码:31 / 35
页数:4
相关论文
共 50 条
  • [1] Estimation of Reference Gamma Radiation Field for Calibration of Radiation Monitors
    Singh, Sunil K.
    Shaiju, Liji
    Gupta, Aashna
    Tripathi, S. M.
    Sathian, V.
    Chaudhury, Probal
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2024, 39 (01): : 31 - 35
  • [2] CALIBRATION OF RADIATION MONITORS AT THE BUREAU OF STANDARDS
    RAPLEY, LF
    HEALTH PHYSICS, 1985, 49 (02): : 361 - 361
  • [3] Simulation of gamma minitype reference radiation for calibration of personal dosimeter
    Liu, Yi-Xin
    Qian, Yi-Kun
    Chen, Zhi-Qiang
    Wei, Biao
    Feng, Peng
    Mao, Ben-Jiang
    NUCLEAR SCIENCE AND TECHNIQUES, 2020, 31 (01)
  • [4] Simulation of gamma minitype reference radiation for calibration of personal dosimeter
    Yi-Xin Liu
    Yi-Kun Qian
    Zhi-Qiang Chen
    Biao Wei
    Peng Feng
    Ben-Jiang Mao
    NuclearScienceandTechniques, 2020, 31 (01) : 114 - 124
  • [5] Simulation of gamma minitype reference radiation for calibration of personal dosimeter
    Yi-Xin Liu
    Yi-Kun Qian
    Zhi-Qiang Chen
    Biao Wei
    Peng Feng
    Ben-Jiang Mao
    Nuclear Science and Techniques, 2020, 31
  • [6] Automation of the measurement system for calibration in the field of gamma radiation
    Knyziak, Adrian Bozydar
    PRZEGLAD ELEKTROTECHNICZNY, 2011, 87 (04): : 253 - 259
  • [7] CALIBRATION OF COLOR-TELEVISION RADIATION MONITORS
    GOLLNICK, DA
    HEALTH PHYSICS, 1973, 25 (03): : 310 - 310
  • [8] An Automatic Calibration System for Portable Radiation Monitors
    Manuel Garcia-Hernandez, Jose
    Mondragon-Contreras, Luis
    Antonio Torres-Bribiesca, Marco
    Lopez-Callejas, Regulo
    Javier Ramirez-Jimenez, Francisco
    Gutierrez-Ocampo, Efren
    Mario Vazquez-Cervantes, Raul
    Cruz-Estrada, Pedro
    Pena-Eguiluz, Rosendo
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 1537 - 1543
  • [9] THE CALIBRATION OF NEUTRON MONITORS FOR RADIATION PROTECTION PURPOSES AT THE CSIR
    WEIDEMA, O
    HEALTH PHYSICS, 1986, 50 (02): : 316 - 316
  • [10] Certification of a Specialized Source of Gamma Radiation Spatially Combined with a Model Reference Neutron Radiation Field
    A. A. Gromov
    A. P. Zhanzhora
    V. F. Shikalov
    Measurement Techniques, 2021, 64 : 244 - 249