Electromagnetic interference of implantable cardiac devices from a shoulder massage machine

被引:0
作者
Saeko Yoshida
Kousaku Fujiwara
Satoshi Kohira
Minoru Hirose
机构
[1] Kitasato University,Department of Medical Safety Engineering, Graduate School of Medical Science
[2] Kitasato University,Clinical Engineering Course, School of Allied Health Science
[3] Kitasato University Hospital,Department of Medical Engineering
来源
Journal of Artificial Organs | 2014年 / 17卷
关键词
Pacemaker; EMI; Massage machine;
D O I
暂无
中图分类号
学科分类号
摘要
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter–defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a “don’t use” warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
引用
收藏
页码:243 / 249
页数:6
相关论文
共 17 条
[1]  
Hirose M(2005)Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven PACE. 28 540-548
[2]  
Hida M(1998)The effect of electromagnetic interference in pacemaker implanted patients by cellular phones J Arrhythm. 14 228-294
[3]  
Sato E(1996)Electromagnetic interference of pacemaker by mobile phones PACE. 19 1431-1446
[4]  
Yamamoto T(2007)Electronic article surveillance systems and interactions with implantable cardiac devices: risk of adverse interactions in public and commercial spaces Mayo Clin Proc. 82 318-322
[5]  
Matsumoto K(1984)Interference in pacemaker PACE. 7 1021-1048
[6]  
Saitou J(2006)Paradoxical atrial undersensing: noise rate reversion or amplifier ringing? J Cardiovasc Electrophysiol. 17 1371-1374
[7]  
Irnich W(1996)Electromagnetic interference of implantable cardiac pacemaker by portable telephones Card Pacing 12 488-497
[8]  
Batz L(undefined)undefined undefined undefined undefined-undefined
[9]  
Muller R(undefined)undefined undefined undefined undefined-undefined
[10]  
Gimble JR(undefined)undefined undefined undefined undefined-undefined