Considerations in Selecting a Fitting a Function for Film Dosimetry in Radiation Therapy

被引:0
|
作者
Ui-Jung Hwang
Ki Hwan Kim
Young Kyung Lim
Haksoo Kim
Sang Hyoun Choi
Byung Jun Min
Hojin Jeong
Jeong Eun Rah
机构
[1] Chungnam National University Hospital,Department of Radiation Oncology
[2] National Cancer Center,Proton Therapy Center
[3] Korea Institute of Radiological & Medical Sciences,Research Center for Radiotherapy
[4] Chungbuk National University Hospital,Department of Radiation Oncology
[5] Gyeongsang National University,Department of Radiation Oncology, Institute of Health Sciences
[6] Myongji Hospital,Department of Radiation Oncology
来源
关键词
Radiation therapy; Quality assurance; EBT3 film; OD to dose calibration; Curve fitting;
D O I
暂无
中图分类号
学科分类号
摘要
Although many technologies, such as an electronic portal imaging device, a software-based quality assurance system, an ion chamber array, etc., have been developed for the purpose of quality assurance of radiotherapy planning, radiochromic film is still one of the most reliable technologies and is being widely used. In particular, compared with other technologies, EBT3 film has unique advantages, including its unlimited spatial resolution, relatively small volume and waterproof quality, which means it can be used inside the patient’s body. The main obstacle to using EBT3 film is that various equations have to be used in the calibration process, which makes it very sensitive to the results. The manufacturer provides a fitting function, but many software programs and researchers use their own functions. In this study, some functions normally used for calibrating EBT3 films are compared by analyzing a couple of characteristics in the course of fitting the function to the measured values. Although our study shows the lack of a suitable fitting function that works well in every situation, it can provide valuable insight into selecting suitable functions for fitting in a clinical setting.
引用
收藏
页码:94 / 101
页数:7
相关论文
共 50 条
  • [1] Considerations in Selecting a Fitting a Function for Film Dosimetry in Radiation Therapy
    Hwang, Ui-Jung
    Kim, Ki Hwan
    Lim, Young Kyung
    Kim, Haksoo
    Choi, Sang Hyoun
    Min, Byung Jun
    Jeong, Hojin
    Rah, Jeong Eun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2020, 77 (01) : 94 - 101
  • [2] FILM DOSIMETRY IN RADIATION THERAPY
    FEHRENTZ, D
    ZUNTER, F
    STRAHLENTHERAPIE, 1968, 135 (03) : 301 - &
  • [3] Film dosimetry for intensity modulated radiation therapy
    Benites-Rengifo, J
    Martínez-Dávalos, A
    Celis, M
    Lárraga, J
    MEDICAL PHYSICS, 2004, 724 : 244 - 248
  • [4] A filtration method for improving film dosimetry in photon radiation therapy
    Yeo, IJ
    Wang, CKC
    Burch, SE
    MEDICAL PHYSICS, 1997, 24 (12) : 1943 - 1953
  • [5] Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluation
    Ju, SG
    Ahn, YC
    Huh, SJ
    Yeo, IJ
    MEDICAL PHYSICS, 2002, 29 (03) : 351 - 355
  • [6] Absolute film dosimetry for verification in intensity modulated radiation therapy
    Mercay, A
    Moeckli, R
    Pache, G
    Bulling, S
    Do, HP
    Bressan, S
    Valley, JF
    STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 : 108 - 108
  • [7] Stereotactic Body Radiation Therapy (SBRT) II: Physics and Dosimetry Considerations
    Benedict, S.
    Yenice, K.
    MEDICAL PHYSICS, 2008, 35 (06) : 2858 - +
  • [8] SELECTING RADIOGRAPHIC FILM - CONSIDERATIONS FOR THE SHOP
    BOWLES, W
    MAYBERRY, G
    MATERIALS EVALUATION, 1988, 46 (04) : 426 - 429
  • [9] Selecting for radiation therapy
    Poortmans, Philip
    FUTURE ONCOLOGY, 2017, 13 (08) : 9 - 9
  • [10] Challenges in Radiochromic Film Dosimetry Applied to Microbeam Radiation Therapy (MRT)
    Anderson, D.
    Warkentin, B.
    Siegbahn, E. A.
    Fallone, B. G.
    MEDICAL PHYSICS, 2011, 38 (06)